Histologic Evaluation of Blood Vessels Sealed with 1,470-nm Diode Laser: Determination of Adequate Condition for Laser Vessel Sealing

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류의학일반

초록

Introduction Energy-based devices allow for a more rapid and efficient ligation of blood vessels during operations. In the present study, we evaluated the feasibility of a laser as an alternative energy source for the vessel sealing system and determined the optimal condition of laser for an effective vessel sealing through histologic examination. Materials and Methods The arteries (5 mm diameter) harvested from porcine legs were compressed between two glass-slides to eliminate its luminal space and were irradiated with 1,470-nm diode laser under various sealing conditions, including laser power (5-30 W), irradiation time (5 or 10 seconds), and focus mode (focus or defocus). Subsequently, the irradiated vessels were fixed in 4% formaldehyde and then processed to paraffin block. The paraffinized sample was sectioned and stained with hematoxylin and eosin for histological evaluation. Results The extent of tissue change was positively correlated with duration and power of laser. In defocus mode, the irradiated vessels showed sufficient tissue denaturation for sealing effect without severe tissue destruction. Moreover, among the various conditions of irradiation, laser power between 15 and 20 W, as well as exposure time of 5 seconds were appropriate for sealing the blood vessels. Conclusion Adequate power and irradiation duration of laser can render blood vessels to be sealed effectively, although the higher power of laser may be required to cut the vessels.

Abstract

Introduction Energy-based devices allow for a more rapid and efficient ligation of blood vessels during operations. In the present study, we evaluated the feasibility of a laser as an alternative energy source for the vessel sealing system and determined the optimal condition of laser for an effective vessel sealing through histologic examination. Materials and Methods The arteries (5 mm diameter) harvested from porcine legs were compressed between two glass-slides to eliminate its luminal space and were irradiated with 1,470-nm diode laser under various sealing conditions, including laser power (5-30 W), irradiation time (5 or 10 seconds), and focus mode (focus or defocus). Subsequently, the irradiated vessels were fixed in 4% formaldehyde and then processed to paraffin block. The paraffinized sample was sectioned and stained with hematoxylin and eosin for histological evaluation. Results The extent of tissue change was positively correlated with duration and power of laser. In defocus mode, the irradiated vessels showed sufficient tissue denaturation for sealing effect without severe tissue destruction. Moreover, among the various conditions of irradiation, laser power between 15 and 20 W, as well as exposure time of 5 seconds were appropriate for sealing the blood vessels. Conclusion Adequate power and irradiation duration of laser can render blood vessels to be sealed effectively, although the higher power of laser may be required to cut the vessels.

참고문헌 (13)

  1. Kennedy JS, "Highburst-strength, feedback-controlled bipolar vessel sealing", Surg Endosc, 12, pp.876~878 (1998)
  2. Payne JH Jr, "Ultrasonic dissection", Surg Endosc, 8, pp.416~418 (1994)
  3. Person B, "Comparison of four energy-based vascular sealing and cutting instruments: a porcine model", Surg Endosc, 22, pp.534~538 (2008)
  4. Levy B, "Randomized trial of suture versus electrosurgical bipolar vessel sealing in vaginal hysterectomy", Obstet Gynecol, 102, pp.147~151 (2003)
  5. Marcello PW, "Vascular pedicle ligation techniques during laparoscopic colectomy. A prospective randomized trial", Surg Endosc, 20, pp.263~269 (2006)
  6. Romano F, "Laparoscopic splenectomy: ligasure versus EndoGIA: a comparative study", J Laparoendosc Adv Surg Tech A, 17, pp.763~767 (2007)
  7. Kim FJ, "Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT)", Surg Endosc, 22, pp.1464~1469 (2008)
  8. Phillips CK, "Tissue response to surgical energy devices", Urology, 71, pp.744~748 (2008)
  9. Govekar HR, "Residual heat of laparoscopic energy devices: how long must the surgeon wait to touch additional tissue", Surg Endosc, 25, pp.3499~3502 (2011)
  10. Cilip CM, "Infrared laser thermal fusion of blood vessels: preliminary ex vivo tissue studies", J Biomed Opt, 18, pp.58001~ (2013)
  11. Giglio NC, "Rapid sealing and cutting of porcine blood vessels, ex vivo, using a high-power, 1470-nm diode laser", J Biomed Opt, 19, pp.38002~ (2014)
  12. Hope WW, "An evaluation of electrosurgical vesselsealing devices in biliary tract surgery in a porcine model", HPB (Oxford), 12, pp.703~708 (2010)
  13. Newcomb WL, "Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices", Surg Endosc, 23, pp.90~96 (2009)

관련 논문AI추천

혈관결찰기의 성능평가방법으로서 혈관파열압력시험의 유용성 연구강보람, 김광택, 김선태, 박아름, 이동혁대한의용생체공학회 · 의공학회지2020 · 제41권 제5호레이저를 이용한 하지정맥류 치료의 정량화 연구강현욱, 안민우, 오정환, Nguyen Van Phuc대한의용생체공학회 · 의공학회지2014 · 제35권 제3호Diode Laser—Can It Replace the Electrical Current Used in Endoscopic Submucosal Dissection?Bong Min Ko, Gwang Ho Baik, Hee Seok Moon, In Kyung Yoo, Jae-Young Jang, Jin Seok Jang, Joo Young Cho, Seong Hwan Kim, Sun Gyo Lim, Wan-Sik Lee, Weon Jin Ko, Young Kwan Cho, Yunho Jung대한소화기내시경학회 · Clinical Endoscopy2021 · 제54권 제4호Efficacy of Oxidized Regenerated Cellulose, SurgiGuard®, in Porcine Surgery김경식, 김성현, 김세훈, 김현균, 윤혜성연세대학교의과대학 · Yonsei Medical Journal2017 · 제58권 제1호A Comparison of the Effects of Solid-state Lasers on Normal Guinea-pig Muscle and Skin: Using 532 nm Flashlamp-excited and Diode-excited Lasers김종수, 우승훈, 이상준, Gil Jae Jung, Hae Ryung Choi, Phil Sang Chung, Yoon Ah Kim대한의학레이저학회 · Medical Lasers2022 · 제11권 제1호Comparison of Internally Cooled Wet Electrode and Hepatic Vascular Inflow Occlusion Method for Hepatic Radiofrequency Ablation박미현, 전경식, 조준식, Byungmo Lee, Byung Seok Shin, Kichang Lee거트앤리버 소화기연관학회협의회 · Gut and Liver2012 · 제6권 제4호

이 간행물의 다른 논문 (10건)