본문 바로가기
KERD
논문KCI 등재

T1-weighted MR Imaging of the Neonatal Brain at 3.0 Tesla: Comparison of Spin Echo, Fast Inversion Recovery, and Magnetization-prepared Three Dimensional Gradient Echo Techniques

1

조회

0

다운로드

3

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Purpose : The purpose of this study was to evaluate the usefulness of fast inversion recovery (FIR) and magnetization-prepared three dimensional gradient echo sequence (3D GRE) T1-weighted sequences for neonatal brain imaging compared with spin echo (SE) sequence in a 3T MR unit. Materials and Methods : T1-weighted axial SE, FIR and 3D GRE sequences were evaluated from 3T brain MR imaging in 20 neonates. The signal-to-noise ratio (SNR) of different tissues was measured and contrast-to-noise ratios (CNR) were determined and compared in each of the sequences. Visual analysis was carried out by grading gray-white matter differentiation, myelination, and artifacts. The Wilcoxon signed ranked test was used for evaluation of the statistical significance of CNR differences between the sequences. Results : Among the three sequences, the 3D GRE had the best SNRs. CNRs obtained with FIR and 3D GRE were statistically superior to those obtained with SE; these CNRs were better on the 3D GRE compared to the FIR. Gray to white matter differentiation and myelination were better delineated on the FIR and 3D GRE than the SE. However, motion artifacts were more commonly observed on the 3D GRE and flow-related artifacts of vessels were frequently seen on the FIR. Conclusion : FIR and 3D GRE are valuable alternative T1-weighted sequences to conventional SE imaging of the neonatal brain at 3T providing superior image quality.

Abstract

Purpose : The purpose of this study was to evaluate the usefulness of fast inversion recovery (FIR) and magnetization-prepared three dimensional gradient echo sequence (3D GRE) T1-weighted sequences for neonatal brain imaging compared with spin echo (SE) sequence in a 3T MR unit. Materials and Methods : T1-weighted axial SE, FIR and 3D GRE sequences were evaluated from 3T brain MR imaging in 20 neonates. The signal-to-noise ratio (SNR) of different tissues was measured and contrast-to-noise ratios (CNR) were determined and compared in each of the sequences. Visual analysis was carried out by grading gray-white matter differentiation, myelination, and artifacts. The Wilcoxon signed ranked test was used for evaluation of the statistical significance of CNR differences between the sequences. Results : Among the three sequences, the 3D GRE had the best SNRs. CNRs obtained with FIR and 3D GRE were statistically superior to those obtained with SE; these CNRs were better on the 3D GRE compared to the FIR. Gray to white matter differentiation and myelination were better delineated on the FIR and 3D GRE than the SE. However, motion artifacts were more commonly observed on the 3D GRE and flow-related artifacts of vessels were frequently seen on the FIR. Conclusion : FIR and 3D GRE are valuable alternative T1-weighted sequences to conventional SE imaging of the neonatal brain at 3T providing superior image quality.

참고문헌 (15)

  1. "Advanced MRtechniques in the term-born neonate with perinatal braininjury", 10, pp.445~460 (2005)
  2. "MRimaging of the neonatal brain at 3 Tesla", 8, pp.281~289 (2004)
  3. "MR imaging at highmagnetic fields", 46, pp.45~52 (2003)
  4. "Highfield body MR imaging: preliminary experiences", 28, pp.159~162 (2004)
  5. "Usefulness ofT1-weighted image with fast inversion recovery technique inintracranial lesions: comparison with T1-weighted spin echoimage", 24, pp.263~269 (2000)
  6. "RARE imaging: a fastimaging method for clinical MR"
  7. "Jr. NMRrelaxation times in the human brain at 3.0 tesla", 9, pp.531~538 (1999)
  8. "T1-weighted MR imaging of thebrain using a fast inversion recovery pulse sequence", 6, pp.356~362 (1996)
  9. "MR imaging: clinical use of theinversion recovery sequence"
  10. "Three-dimensional, T1-weighted gradient-echo imaging ofthe brain with a volumetric interpolated examination", 23, pp.995~1002 (2002)
  11. "T1-weighted imaging of the brain at 3 teslausing a 2-dimensional spoiled gradient echo technique", 41, pp.68~75 (2006)
  12. "Efficacy of contrast medium use for neuroimaging at 3.0T: utility of IR-FSE compared to other T1-weighted pulsesequences", 29, pp.499~505 (2005)
  13. "Comparison of a T1-weighted inversion-recovery-, gradient-echo- and spin-echo sequence for imagingof the brain at 3.0 Tesla", 177(4), pp.536~542 (2005)
  14. "T1-weighted fluid-attenuated inversion recovery at lowfield strength: a viable alternative for T1-weighted intracranialimaging", 24, pp.648~651 (2003)
  15. "Enhancing gray-to-white matter contrast in 3T T1 spin-echo brain scans by optimizing flip angle", 26(8), pp.2000~2004 (2005)

관련 논문AI추천

MRI Findings to Predict Neurodevelopmental Outcomes in Preterm Infants Near Term-Equivalent Age김성신, 박가영, 홍현숙대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2020 · 제24권 제1호Implementation of the Feed and Swaddle Technique as a Non-Pharmacological Strategy to Conduct Brain Magnetic Resonance Imaging in Very Low Birth Weight Infants유영명, Jang Hoon Lee, Ji Eun Park, Moon Sung Park대한신생아학회 · Neonatal medicine2021 · 제28권 제3호T1-weighted FLAIR MR Imaging for the Evaluation of Enhancing Brain Tumors: Comparison with Spin Echo Imaging나재범, 박미정, 신화선, 전경녀, 정보슬, 정성훈, 최대섭, 최혜영대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2014 · 제18권 제2호신생아 중환자실을 퇴원하는 고위험 환아에서 순차적인 뇌초음파 검사 후 시행한 자기 공명 영상의 유용성김지혜대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2008 · 제12권 제2호미숙아에서 뇌 MRI와 뇌 초음파 검사의 결과 비교김병일, 김이경, 김한석, 손진아, 이승현, 이주영, 이진아, 최중환, 최창원대한신생아학회 · Neonatal medicine2012 · 제19권 제1호초극소 저체중 출생아의 뇌 MRI 상 Maturation Score의 의의김규리, 김병일, 김수영, 김우선, 김윤주, 김이경, 김인원, 김지영, 김한석, 손진아, 송인규, 신승한, 이승현, 이재명, 이주영, 이진아, 천정은, 최중환, 최창원대한신생아학회 · Neonatal medicine2011 · 제18권 제2호

이 간행물의 다른 논문 (5건)