Radiation Exposure of the Hand and Chest during C-arm Fluoroscopy-Guided Procedures

2

조회

0

다운로드

9

인용

0

참고문헌

키워드

초록

Background:The C-arm fluoroscope is an essential tool for the intervention of pain. The aim of this study was to investigate the radiation exposure experienced by the hand and chest of pain physicians during C-arm fluoroscopy-guided procedures. Methods:This is a prospective study about radiation exposure to physicians during transforaminal epidural steroid injection (TFESI) and medial branch block (MBB). Four pain physicians were involved in this study. Data about effective dose (ED) at each physician’s right hand and left side of the chest, exposure time, radiation absorbed dose (RAD), and the distance from the center of the X-ray field to the physician during X-ray scanning were collected. Results:Three hundred and fifteen cases were included for this study. Demographic data showed no significant differences among the physicians in the TFESIs and MBBs. In the TFESI group, there was a significant difference between the ED at the hand and chest in all the physicians. In physician A, B and C, the ED at the chest was more than the ED at the hand. The distance from the center of the X-ray field to physician A was more than that of the other physicians, and for the exposure time, the ED and RAD in physician A was less than that of the other physicians. In the MBB group, there was no difference in the ED at the hand and chest, except for physician D. The distance from the center of the X-ray field to physician A was more than that of the other physicians and the exposure time in physician A was less than that of the other physicians. Conclusions:In conclusion, the distance from the radiation source, position of the hand, experience and technique can correlate with the radiation dose.

Abstract

Background:The C-arm fluoroscope is an essential tool for the intervention of pain. The aim of this study was to investigate the radiation exposure experienced by the hand and chest of pain physicians during C-arm fluoroscopy-guided procedures. Methods:This is a prospective study about radiation exposure to physicians during transforaminal epidural steroid injection (TFESI) and medial branch block (MBB). Four pain physicians were involved in this study. Data about effective dose (ED) at each physician’s right hand and left side of the chest, exposure time, radiation absorbed dose (RAD), and the distance from the center of the X-ray field to the physician during X-ray scanning were collected. Results:Three hundred and fifteen cases were included for this study. Demographic data showed no significant differences among the physicians in the TFESIs and MBBs. In the TFESI group, there was a significant difference between the ED at the hand and chest in all the physicians. In physician A, B and C, the ED at the chest was more than the ED at the hand. The distance from the center of the X-ray field to physician A was more than that of the other physicians, and for the exposure time, the ED and RAD in physician A was less than that of the other physicians. In the MBB group, there was no difference in the ED at the hand and chest, except for physician D. The distance from the center of the X-ray field to physician A was more than that of the other physicians and the exposure time in physician A was less than that of the other physicians. Conclusions:In conclusion, the distance from the radiation source, position of the hand, experience and technique can correlate with the radiation dose.

참고문헌 (17)

  1. Andreassi MG, "The biological effects of diagnostic cardiac imaging on chronically exposed physicians: the importance of being non-ionizing", Cardiovasc Ultrasound, 2, pp.25~ (2004)
  2. Pyong Eun Park, "Radiation Safety and Education in the Applicants of the Final Test for the Expert of Pain Medicine", The Korean Journal of Pain, 25(1), pp.16~21 (2012), 대한통증학회
  3. Pradhan AS, "Evolution of dosimetric quantities of International Commission on Radiological Protection (ICRP): Impact of the forthcoming recommendations", J Med Phys, 32, pp.89~91 (2007)
  4. Miller DL, "Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology", J Vasc Interv Radiol, 21, pp.607~615 (2010)
  5. Harding LK, "International commission on radiation protection", Nucl Med Commun, 11, pp.585~587 (1990)
  6. Cousins C, "Medical interventional procedures- reducing the radiation risks", Clin Radiol, 59, pp.468~473 (2004)
  7. Mroz TE, "Radiation exposure to the surgeon and the patient during kyphoplasty", J Spinal Disord Tech, 21, pp.96~100 (2008)
  8. Theocharopoulos N, "Occupational exposure from common fluoroscopic projections used in orthopaedic surgery", J Bone Joint Surg Am, 85, pp.1698~1703 (2003)
  9. Blattert TR, "Skill dependence of radiation exposure for the orthopaedic surgeon during interlocking nailing of long-bone shaft fractures: a clinical study", Arch Orthop Trauma Surg, 124, pp.659~664 (2004)
  10. Bahari S, "Radiation exposure of the hands and thyroid gland during percutaneous wiring of wrist and hand procedures", Acta Orthop Belg, 72, pp.194~198 (2006)
  11. Arnstein PM, "The risk from radiation exposure during operative X-ray screening in hand surgery", J Hand Surg Br, 19, pp.393~396 (1994)
  12. Siegel JA, "Calculating the absorbed dose from radioactive patients: the line-source versus point-source model", J Nucl Med, 43, pp.1241~1244 (2002)
  13. Bushberg JT, "The essential physics of medical imaging. 3rd ed", pp.837~902 (2011), Lippincott Williams & Wilkins
  14. Singer G, "Occupational radiation exposure to the surgeon", J Am Acad Orthop Surg, 13, pp.69~76 (2005)
  15. Jae Hun Cho, "A Study to Compare the Radiation Absorbed Dose of the C-arm Fluoroscopic Modes", The Korean Journal of Pain, 24(4), pp.199~204 (2011), 대한통증학회
  16. Kallmes DF, "Radiation dose to the operator during vertebroplasty: prospective comparison of the use of 1-cc syringes versus an injection device", AJNR Am J Neuroradiol, 24, pp.1257~1260 (2003)
  17. Kruger R, "Radiation dose reduction to medical staff during vertebroplasty: a review of techniques and methods to mitigate occupational dose", Spine, 28, pp.1608~1613 (2003)

이 간행물의 다른 논문 (19건)