Comparison of resorbable plates and titanium plates for fixatio stability of combined mandibular symphysis and angle fractures
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Objectives: We compared resorbable plates with titanium plates for treatment of combined mandibular angle and symphyseal fractures.Materials and Methods: Patients with mandibular angle and symphysis fractures were divided into two groups. The control (T) group received tita-nium plates while the experimental (R) group received resorbable plates. All procedures were carried out under general anesthesia using standard surgi-cal techniques. We compared the frequency of wound dehiscence, development of infection, malocclusion, malunion, screw breakage, and any other technical difficulties between the two groupsResults: Thirteen patients were included in the R group, where 39 resorbable plates were applied. The T group consisted of 16 patients who received 48 titanium plates. The mean age in the R and T groups was 28.29 and 24.23 years, respectively. Primary healing of the fractured mandible was ob-tained in all patients in both groups. Postoperative complications were minor and transient. Moreover, there were no significantdifferences in the rates of various complications between the two groups. Breakage of 3 screws during the perioperative period was seen in the R group, while no screws or plates were broken in the T group.Conclusion: Resorbable plates can be used to stabilize combined mandibular angle and symphysis fractures.
Abstract
Objectives: We compared resorbable plates with titanium plates for treatment of combined mandibular angle and symphyseal fractures.Materials and Methods: Patients with mandibular angle and symphysis fractures were divided into two groups. The control (T) group received tita-nium plates while the experimental (R) group received resorbable plates. All procedures were carried out under general anesthesia using standard surgi-cal techniques. We compared the frequency of wound dehiscence, development of infection, malocclusion, malunion, screw breakage, and any other technical difficulties between the two groupsResults: Thirteen patients were included in the R group, where 39 resorbable plates were applied. The T group consisted of 16 patients who received 48 titanium plates. The mean age in the R and T groups was 28.29 and 24.23 years, respectively. Primary healing of the fractured mandible was ob-tained in all patients in both groups. Postoperative complications were minor and transient. Moreover, there were no significantdifferences in the rates of various complications between the two groups. Breakage of 3 screws during the perioperative period was seen in the R group, while no screws or plates were broken in the T group.Conclusion: Resorbable plates can be used to stabilize combined mandibular angle and symphysis fractures.
참고문헌 (20)
- Ellis E 3rd, "Ten years of mandibular fractures: an analysis of 2,137 cases", Oral Surg Oral Med Oral Pathol, 59, pp.120~129 (1985)
- Haug RH, "An epidemiologic survey of facial fractures and concomitant injuries", J Oral Maxillofac Surg, 48, pp.926~932 (1990)
- Tuovinen V, "A retrospective analysis of 279 patients with isolated mandibular fractures treated with titanium miniplates", J Oral Maxillofac Surg, 52, pp.931~935 (1994)
- Kim YK, "Tissue response to titanium plates: a transmitted electron microscopic study", J Oral Maxillofac Surg, 55, pp.322~326 (1997)
- Suuronen R, "Bioabsorbable plates and screws: current state of the art in facial fracture repair", J Craniomaxillofac Trauma, 6, pp.19~27 (2000)
- Leonhardt H, "INION compared with titanium osteosynthesis: a prospective investigation of the treatment of mandibular fractures", Br J Oral Maxillofac Surg, 46, pp.631~634 (2008)
- 최은주, "Monocortical Osteosynthesis 이론에 따른 하악골 우각부 골절 수술시 Titanium miniplate와 Biodegradable miniplate의 임상적, 방사선학적 비교 연구", 대한구강악안면외과학회지, 32(3), pp.222~225 (2006), 대한구강악안면외과학회
- 정종철, "악안면 골절 치료시 흡수성 고정판의 사용에 관한 임상 연구", 대한구강악안면외과학회지, 29(6), pp.438~443 (2003), 대한구강악안면외과학회
- Kim YK, "Treatment of mandible fractures using bioabsorbable plates", Plast Reconstr Surg, 110, pp.25~31 (2002)
- Champy M, "Mandibular osteosynthesis by miniature screwed plates via a buccal approach", J Maxillofac Surg, 6, pp.14~21 (1978)
- Bos RR, "Bio-absorbable plates and screws for internal fixation of mandibular fractures. A study of six dogs", Int J Oral Maxillofac Surg, 18, pp.365~369 (1989)
- 김영균, "Application of Bioabsorbable Plates in Orthognathic Surgery", 대한구강악안면외과학회지, 32(1), pp.60~64 (2006), 대한구강악안면외과학회
- Bessho K, "A bioabsorbable poly-L-lactide miniplate and screw system for osteosynthesis in oral and maxillofacial surgery", J Oral Maxillofac Surg, 55, pp.941~945 (1997)
- Kallela I, "Fixation of mandibular body osteotomies using biodegradable amorphous self-reinforced (70L:30DL) polylactide or metal lag screws: an experimental study in sheep", J Craniomaxillofac Surg, 27, pp.124~133 (1999)
- Wittwer G, "Complications after zygoma fracture fixation: is there a difference between biodegradable materials and how do they compare with titanium osteosynthesis?", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 101, pp.419~425 (2006)
- Bayram B, "Comparison of fixation stability of resorbable versus titanium plate and screws in mandibular angle fractures", J Oral Maxillofac Surg, 67, pp.1644~1648 (2009)
- Esen A, "Comparison of stability of titanium and absorbable plate and screw fixation for mandibular angle fractures", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 106, pp.806~811 (2008)
- Ellis E 3rd, "Open reduction and internal fixation of combined angle and body", symphysis fractures of the mandible: how much fixation is enough?, 2013, pp.71, 726~733
- Fonseca RJ, "Oral and Maxillofacial Trauma. Vol 2" (1997), WB saunders
- Ahmed W, "Bioresorbable versus titanium plates for mandibular fractures", J Coll Physicians Surg Pak, 23, pp.480~483 (2013)