Awareness during general anesthesia despite simultaneous bispectral index and end-tidal anesthetic gas concentration monitoring
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Awareness during general anesthesia occurs in approximately 0.1–0.2% of cases; nevertheless, particular attention is required because it can lead to critical complications including insomnia, depression, anxiety, and post-traumatic stress disorder. To prevent these complications, bispectral index (BIS) and end-tidal anesthetic gas (ETAG) concentration monitoring are commonly used to examine patient consciousness during surgery. In the present case, an 80-year-old man was scheduled for total gastrectomy. Anesthesia was maintained using desflurane 4.0–5.0% vol, oxygen, and nitrous oxide. The authors simultaneously monitored BIS, which was maintained between 37 and 43, and ETAG, which was maintained between 0.9 and 1.2 minimum alveolar concentration (MAC). After the operation, however, the authors were surprised to learn that the patient complained of awareness during anesthesia. Although BIS and ETAG concentration monitoring are useful in preventing awareness during anesthesia, they cannot be completely trusted. Even though BIS was maintained at approximately 40 and ETAG at 0.7–1.3 MAC, awareness during anesthesia occurred.
Abstract
Awareness during general anesthesia occurs in approximately 0.1–0.2% of cases; nevertheless, particular attention is required because it can lead to critical complications including insomnia, depression, anxiety, and post-traumatic stress disorder. To prevent these complications, bispectral index (BIS) and end-tidal anesthetic gas (ETAG) concentration monitoring are commonly used to examine patient consciousness during surgery. In the present case, an 80-year-old man was scheduled for total gastrectomy. Anesthesia was maintained using desflurane 4.0–5.0% vol, oxygen, and nitrous oxide. The authors simultaneously monitored BIS, which was maintained between 37 and 43, and ETAG, which was maintained between 0.9 and 1.2 minimum alveolar concentration (MAC). After the operation, however, the authors were surprised to learn that the patient complained of awareness during anesthesia. Although BIS and ETAG concentration monitoring are useful in preventing awareness during anesthesia, they cannot be completely trusted. Even though BIS was maintained at approximately 40 and ETAG at 0.7–1.3 MAC, awareness during anesthesia occurred.
참고문헌 (14)
- Ghoneim MM, "Awareness during anesthesia", Anesthesiology, 92, pp.597~602 (2000)
- Sebel PS, "The incidence of awareness during anesthesia: a multicenter United States study", Anesth Analg, 99, pp.833~839 (2004)
- Lennmarken C, "Victims of awareness", Acta Anaesthesiol Scand, 46, pp.229~231 (2002)
- Avidan MS, "Prevention of intraoperative awareness in a high-risk surgical population", N Engl J Med, 365, pp.591~600 (2011)
- Pollard RJ, "Intraoperative awareness in a regional medical system: a review of 3 years' data", Anesthesiology, 106, pp.269~274 (2007)
- Errando CL, "Awareness with recall during general anaesthesia: a prospective observational evaluation of 4001 patients", Br J Anaesth, 101, pp.178~185 (2008)
- Pandit JJ, "The 5th National Audit Project (NAP5) on accidental awareness during general anaesthesia: summary of main findings and risk factors", Anaesthesia, 69, pp.1089~1101 (2014)
- Song D, "Titration of volatile anesthetics using bispectral index facilitates recovery after ambulatory anesthesia", Anesthesiology, 87, pp.842~848 (1997)
- Rampil IJ, "A primer for EEG signal processing in anesthesia", Anesthesiology, 89, pp.980~1002 (1998)
- Dahaba AA, "Different conditions that could result in the bispectral index indicating an incorrect hypnotic state", Anesth Analg, 101, pp.765~773 (2005)
- Schneider G, "Bispectral Index (BIS) may not predict awareness reaction to intubation in surgical patients", J Neurosurg Anesthesiol, 14, pp.7~11 (2002)
- Myles PS, "Bispectral index monitoring to prevent awareness during anaesthesia: the B-Aware randomised controlled trial", Lancet, 363, pp.1757~1763 (2004)
- McLeskey CH, "Awareness during anesthesia", Can J Anaesth, 46, pp.R80~R87 (1999)
- Gonsowski CT, "Subanesthetic concentrations of desflurane and isoflurane suppress explicit and implicit learning", Anesth Analg, 80, pp.568~572 (1995)