본문 바로가기
KERD
논문KCI 등재

The effect of undersizing and tapping on bone to implant contact and implant primary stability: A histomorphometric study on bovine ribs

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

PURPOSE. Implant site preparation may be adjusted to achieve the maximum possible primary stability. The aim of this investigation was to study the relation among bone-to-implant contact at insertion, bone density, and implant primary stability intra-operatively measured by a torque-measuring implant motor, when implant sites were undersized or tapped. MATERIALS AND METHODS. Undersized (n=14), standard (n=13), and tapped (n=13) implant sites were prepared on 9 segments of bovine ribs. After measuring bone density using the implant motor, 40 implants were placed, and their primary stability assessed by measuring the integral of the torque-depth insertion curve. Bovine ribs were then processed histologically, the bone-toimplant contact measured and statistically correlated to bone density and the integral. RESULTS. Bone-toimplant contact and the integral of the torque-depth curve were significantly greater for undersized sites than tapped sites. Moreover, a correlation between bone to implant contact, the integral and bone density was found under all preparation conditions. The slope of the bone-to-implant/density and integral/density lines was significantly greater for undersized sites, while those corresponding to standard prepared and tapped sites did not differ significantly. CONCLUSION. The integral of the torque-depth curve provided reliable information about bone-to-implant contact and primary implant stability even in tapped or undersized sites. The linear relations found among the parameters suggests a connection between extent and modality of undersizing and the corresponding increase of the integral and, consequently, of primary stability. These results might help the physician determine the extent of undersizing needed to achieve the proper implant primary stability, according to the planned loading protocol.

Abstract

PURPOSE. Implant site preparation may be adjusted to achieve the maximum possible primary stability. The aim of this investigation was to study the relation among bone-to-implant contact at insertion, bone density, and implant primary stability intra-operatively measured by a torque-measuring implant motor, when implant sites were undersized or tapped. MATERIALS AND METHODS. Undersized (n=14), standard (n=13), and tapped (n=13) implant sites were prepared on 9 segments of bovine ribs. After measuring bone density using the implant motor, 40 implants were placed, and their primary stability assessed by measuring the integral of the torque-depth insertion curve. Bovine ribs were then processed histologically, the bone-toimplant contact measured and statistically correlated to bone density and the integral. RESULTS. Bone-toimplant contact and the integral of the torque-depth curve were significantly greater for undersized sites than tapped sites. Moreover, a correlation between bone to implant contact, the integral and bone density was found under all preparation conditions. The slope of the bone-to-implant/density and integral/density lines was significantly greater for undersized sites, while those corresponding to standard prepared and tapped sites did not differ significantly. CONCLUSION. The integral of the torque-depth curve provided reliable information about bone-to-implant contact and primary implant stability even in tapped or undersized sites. The linear relations found among the parameters suggests a connection between extent and modality of undersizing and the corresponding increase of the integral and, consequently, of primary stability. These results might help the physician determine the extent of undersizing needed to achieve the proper implant primary stability, according to the planned loading protocol.

참고문헌 (23)

  1. Glauser R, "Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants", Clin Oral Implants Res, 15, pp.428~434 (2004)[DOI]
  2. Mohlhenrich SC, "Effects on primary stability of three different techniques for implant site preparation in synthetic bone models of different densities", Br J Oral Maxillofac Surg, 54, pp.980~986 (2016)[DOI]
  3. Friberg B, "Branemark implants and osteoporosis: a clinical exploratory study", Clin Implant Dent Relat Res, 3, pp.50~56 (2001)[DOI]
  4. Stocchero M, "Biomechanical, biologic, and clinical outcomes of undersized implant surgical preparation: A systematic review", Int J Oral Maxillofac Implants, 31, pp.1247~1263 (2016)[DOI]
  5. Tabassum A, "Bone particles and the undersized surgical technique", J Dent Res, 89, pp.581~586 (2010)[DOI]
  6. Shalabi MM, "Histological evaluation of oral implants inserted with different surgical techniques into the trabecular bone of goats", Clin Oral Implants Res, 18, pp.489~495 (2007)[DOI]
  7. Tabassum A, "Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study", Clin Oral Implants Res, 20, pp.327~332 (2009)[DOI]
  8. Al-Marshood MM, "Study of the osseointegration of dental implants placed with an adapted surgical technique", Clin Oral Implants Res, 22, pp.753~759 (2011)[DOI]
  9. Alghamdi H, "Undersized implant site preparation to enhance primary implant stability in poor bone density: a prospective clinical study", J Oral Maxillofac Surg, 69, pp.506~512 (2011)[DOI]
  10. Danilo Alessio Di Stefano, "A torque-measuring micromotor provides operator independent measurements marking four different density areas in maxillae", The Journal of Advanced of Prosthodontics, 7(1), pp.51~55 (2015), 대한치과보철학회[DOI]
  11. Iezzi G, "Correlation between the bone density recorded by a computerized implant motor and by a histomorphometric analysis: a preliminary in vitro study on bovine ribs", Clin Implant Dent Relat Res, 17, pp.35~44 (2015)[DOI]
  12. Di Stefano DA, "Correlation between bone density and instantaneous torque at implant site preparation: A validation on polyurethane foam blocks of a device assessing density of jawbones", Int J Oral Maxillofac Implants, 31, pp.e128~e135 (2016)
  13. Degidi M, "Development of a new implant primary stability parameter: insertion torque revisited", Clin Implant Dent Relat Res, 15, pp.637~644 (2013)
  14. Iezzi G, "A site-specific intraoperative measurement of bone-to-implant contact during implant insertion: A study on bovine ribs using a computerized implant motor", J Dent Sci, 10, pp.21~27 (2015)[DOI]
  15. Cappare P, "Correlation between initial BIC and the insertion torque", depth integral recorded with an instantaneous torque-measuring implant motor: An in vivo study, 2015, pp.17, e613~e620
  16. Di Stefano DA, "The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks", J Prosthet Dent (2017)
  17. Wang TM, "The effect of implant design and bone quality on insertion torque, resonance frequency analysis, and insertion energy during implant placement in low or low- to medium-density bone", Int J Prosthodont, 28, pp.40~47 (2015)[DOI]
  18. Degidi M, "Influence of underpreparation on primary stability of implants inserted in poor quality bone sites: an in vitro study", J Oral Maxillofac Surg, 73, pp.1084~1088 (2015)[DOI]
  19. Piattelli A, "High-precision, cost-effective cutting system for producing thin sections of oral tissues containing dental implants", Biomaterials, 18, pp.577~579 (1997)[DOI]
  20. Andrade JM, "Statistical comparison of the slopes of two regression lines: A tutorial", Anal Chim Acta, 838, pp.1~12 (2014)[DOI]
  21. Ku HH, "Notes on the use of propagation of error formulas", J Res Natl Bureau Stand, 70C, pp.263~273 (1966)
  22. Fernandes Ede L, "Primary stability of turned and acid-etched screw-type implants: a removal torque and histomorphometric study in rabbits", Int J Oral Maxillofac Implants, 22, pp.886~892 (2007)
  23. Tabassum A, "Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study", Clin Oral Implants Res, 21, pp.213~220 (2010)[DOI]

이 간행물의 다른 논문 (11건)