전극이 전정에 삽입된 인공와우이식술 1예

Misimplantation of an Electrode Array Into the Vestibule in Cochlear Implantation

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류이비인후과학

초록

Surgical complication related to an electrode array in cochlear implantation is rare; however, in cases with cochlear ossification after bacterial meningitis, the electrode array insertion can be problematic. The patient complained of an aggravated hearing impairment after bacterial meningitis. He was performed cochlear implantation. Both intraoperative neural response telemetry (NRT) and impedance telemetry results were normal. However, postoperative transorbital x-ray revealed only partial insertion of the electrode array. Four days after the surgery, he complained dizziness. Postoperative temporal bone CT showed malposition of the electrode array into the vestibule. Revision surgery was performed and a new electrode array was fully inserted after further drilling out into the cochleostomy window. Though intraoperative NRT and impedance telemetry are normal, misdirection of an electrode array can occur. Therefore, it should be considered that intraoperative radiologic examination is used after insertion of an electrode array for assessment of an electrode array location.

Abstract

Surgical complication related to an electrode array in cochlear implantation is rare; however, in cases with cochlear ossification after bacterial meningitis, the electrode array insertion can be problematic. The patient complained of an aggravated hearing impairment after bacterial meningitis. He was performed cochlear implantation. Both intraoperative neural response telemetry (NRT) and impedance telemetry results were normal. However, postoperative transorbital x-ray revealed only partial insertion of the electrode array. Four days after the surgery, he complained dizziness. Postoperative temporal bone CT showed malposition of the electrode array into the vestibule. Revision surgery was performed and a new electrode array was fully inserted after further drilling out into the cochleostomy window. Though intraoperative NRT and impedance telemetry are normal, misdirection of an electrode array can occur. Therefore, it should be considered that intraoperative radiologic examination is used after insertion of an electrode array for assessment of an electrode array location.

참고문헌 (11)

  1. Migirov L, "Revision surgeries in cochlear implant patients: a review of 45 cases", Eur Arch Otorhinolaryngol, 264(1), pp.3~7 (2007)
  2. Bhatia K, "Surgical complications and their management in a series of 300 consecutive pediatric cochlear implantations", Otol Neurotol, 25(5), pp.730~739 (2004)
  3. Rotteveel LJ, "Cochlear implantation in 53 patients with otosclerosis: demographics, computed tomographic scanning, surgery and complications", Otol- Neurotol, 25(6), pp.943~952 (2004)
  4. Tange RA, "Intracochlear misdirected implantation of a cochlear implant", Acta Otolaryngol, 126(6), pp.650~652 (2006)
  5. Ganz BJ, "Use of multichannel cochlear implants in obstructed and obliterated cochleas", Otolaryngol Head Neck Surg, 98(1), pp.72~81 (1988)
  6. Steenerson RL, "Scala vestibuli cochlear implantation for labyrinthine ossification", Am J Otol, 11(5), pp.360~363 (1990)
  7. Lenarz T, "The nucleus double array cochlear implant: a newconcept for the obliterated cochlea", Otol- Neurotol, 22(1), pp.24~32 (2001)
  8. Pau H, "Displacement of electrodes of a cochlear implant into the vestibular system: intra- and postoperative electrophysiological analyses", ActaOtolaryngol, 125(10), pp.1116~1118 (2005)
  9. Carelsen B, "Cochlear implant electrode array insertionmonitoring with intra-operative 3D rotational X-ray", Clin Otolaryngol, 32(1), pp.46~50 (2007)
  10. Böhmer A, "The subjective visual vertical as a clinical parameter of vestibular function in peripheral vestibular diseases", J Vestib Res, 5(1), pp.35~45 (1995)
  11. Rosenhall U, "Auditory brain stem responses in patients with vertigo", Clin Otolaryngol Allied Sci, 9(3), pp.149~154 (1984)

관련 논문AI추천

이 간행물의 다른 논문 (18건)