직장암 3문조사와 자궁경부암 4문조사 방사선 치료 시 유리선량계를 이용한 피부선량 측정에 관한 연구

A Study on the Skin Dose Measurement of Glass Dosimeter Use During Rectal Cancer 3field Technic and Cervical Cancer 4field Technic Radiation Therapy

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Recently a variety of high technologies for radiation therapy (IMRT, SRS,. 3D-RT, etc.) has been developed. For the cervical and rectal cancer, 3field or 4field radiotherapy have been applied to the patients. In the case of two-dimensional treatment, one of the most typical side-effects is skin burn due to the radiation irradiation. In general the skin dose is evaluated by only a single measurement during the whole treatment period. In this study, however, skin dose was measured in each radiation treatment and the total skin dose was accumulated in a glass dosimeter through all the cases. After simulating the skin dose from treatment planning system, the results were compared with the actual skin doses. The results showed a good agreement between two data sets. Even though there are certain amount of errors caused by the patient movement along the treatment, the difference between actual dose and simulated dose was within the accepted range of error.

Abstract

Recently a variety of high technologies for radiation therapy (IMRT, SRS,. 3D-RT, etc.) has been developed. For the cervical and rectal cancer, 3field or 4field radiotherapy have been applied to the patients. In the case of two-dimensional treatment, one of the most typical side-effects is skin burn due to the radiation irradiation. In general the skin dose is evaluated by only a single measurement during the whole treatment period. In this study, however, skin dose was measured in each radiation treatment and the total skin dose was accumulated in a glass dosimeter through all the cases. After simulating the skin dose from treatment planning system, the results were compared with the actual skin doses. The results showed a good agreement between two data sets. Even though there are certain amount of errors caused by the patient movement along the treatment, the difference between actual dose and simulated dose was within the accepted range of error.

참고문헌 (16)

  1. Edward C. Halperin, "Principles and Practice of Radiation Oncology", pp.320~332 (2008)
  2. Jiade J. Lu, "Radiation Oncology An Evidence-Based Approach", pp.245~369 (2008)
  3. Tatsuya Ishikawa, "The basic characteristics of the glass dosimeter for measuring personal dose (II)", Radioisotope journal, 9(3), pp.45~47 (1994)
  4. 라정은, "유리선량계를 이용한 감마나이프의 출력인자 결정", 의학물리, 18(1), pp.13~19 (2007), 한국의학물리학회
  5. 라정은, "유리선량계의 선량 특성에 관한 연구", 방사선방어학회지, 31(4), pp.181~186 (2006), 대한방사선방어학회
  6. 고영은, "유리선량계와 다이오드 측정기를 이용한 유방암 환자의 체표면 선량측정 비교", 의학물리, 19(1), pp.9~13 (2008), 한국의학물리학회
  7. 제재용, "치료 보조기구 사용 시 후 방향 피부선량 측정", 방사선기술과학, 32(1), pp.107~110 (2009), 대한방사선과학회
  8. 박승우, "유리선량계를 이용한 감마선 조사장치의 공간선량분포 측정", 2009년도 한국의학 물리학회 춘계 학술발표회 논문집, pp.106~107 (2009)
  9. 손기홍, "유리선량계의 고선량 직선성, 누적선량, 에너지의존성 특성 평가", 2009년도 한국의학물리학회 추계 학술발표회 논문집, pp.66~67 (2009)
  10. Ho Jun Lee, "Dose Effect of Tissue Compensator for 6 MV X-Ray", J Korean Soc Ther Radiol, 10(2), pp.147~153 (1992)
  11. 김연실, "Comparison of Virtual Wedge versus Physical Wedge Affecting onDose Distribution of Treated Breast and Adjacent Normal Tissuefor Tangential Breast Irradiation", Radiation oncology journal, 22(3), pp.225~233 (2004), 대한방사선종양학회
  12. 김성철, "MS EXCEL 및 Bit system을 이용한 피부선량 계산 프로그램의 제작 및 응용", 방사선기술과학, 32(1), pp.39~43 (2009), 대한방사선과학회
  13. 박철우, "전자선에너지에 따른 치료보조기구의 표면선량 평가", 한국콘텐츠학회 논문지, 10(6), pp.360~363 (2010), 한국콘텐츠학회
  14. 최병기, "방사선 치료시 피부선량 측정 및 평가", 방사선치료기학술지, 11(1), pp.69~72 (1999)
  15. 안승권, "BOLUS 사용시 환자와의 AIR GAP에 의한 피부선량 변화에 관한 고찰", 방치기학술지, 13(1), pp.51~58 (2001)
  16. 김성규, "유방암환자의 세기조절방사선치료에서 피부선량 평가", 의학물리, 18(3), pp.167~171 (2007), 한국의학물리학회

이 간행물의 다른 논문 (8건)