In this study, we have researched the effectiveness of silicone pad. A distribution of scatter ray in mammography was evaluated using Monte-Carlo (MC) simulation technique and then a silicone pad was applied to remove the scatter ray for improving image quality. Molybdenum target and Molybdenum filter combination made a difference of 59.8% to a number of photon at 17.5 keV. On the other hand, Tungsten target and Rhodium filter showed a variation of 24.5% at 20 keV. Mean 68 of SNR was increased in Selenia and mean 1.04 of SNR was raised in Senographe. Silicone pad was significantly effective to reduce the scatter ray that was generated by primary X-ray. It can decrease an absorption rate of scatter ray to patient body and whilst it improve the image quality from increasing SNR.

Abstract

In this study, we have researched the effectiveness of silicone pad. A distribution of scatter ray in mammography was evaluated using Monte-Carlo (MC) simulation technique and then a silicone pad was applied to remove the scatter ray for improving image quality. Molybdenum target and Molybdenum filter combination made a difference of 59.8% to a number of photon at 17.5 keV. On the other hand, Tungsten target and Rhodium filter showed a variation of 24.5% at 20 keV. Mean 68 of SNR was increased in Selenia and mean 1.04 of SNR was raised in Senographe. Silicone pad was significantly effective to reduce the scatter ray that was generated by primary X-ray. It can decrease an absorption rate of scatter ray to patient body and whilst it improve the image quality from increasing SNR.

참고문헌 (15)

  1. Kump K, "Digital X-ray detector technology", Physics in Medicine & Biology, 26(5), pp.907~911 (1981)
  2. Van der Stelt PF, "Better imaging : The advantages of digital radiography", J Am Dent Asso, 139, pp.7S~13S (2008)[DOI]
  3. Carroll QB, "Digital imaging processing" (2014), Charles C Tomas
  4. Ingrid Helen Ryste Hauge, "Digital Mammography", pp.153~161
  5. Amanda Demetri-Lewis, "Breast Calcifications: The Focal Group", American Journal of Roentgenology, 198(4), pp.W325~W343 (2012), American Roentgen Ray Society[DOI]
  6. Pouldnioswki G, "SpekCalc:A program to calculate photon spectra from tungsten ano de x-ray tubes", Phys Med Biol, 54(19), pp.N433~N438 (2009)
  7. Jan S, "GATE - Geant 4 Application for Tomographic Emission: A simulation toolkit for PET and SPECT", Phys Med Biol, 49(19), pp.4543~4561 (2004)
  8. Arefan D, "Calculation of the contrast of the calcification in digital mammography system : Gate validation", J Cancer Res Ther, 14(2), pp.335~340 (2018)
  9. Leon SM, "Characterization of scatter in digital mammography from physical measurements", Med Phys, 41(6), pp.061901~ (2014)
  10. Shrimpton PC, "Electron density values of various human tissues : in vitro Compton scatter measurements andcalculated ranges", Physics in Medicine & Biology, 26(5), pp.907~911 (1981)
  11. 박일, "몬테카를로 방법론을 이용한 구내촬영 검사 시 환자 방사선량 평가", 방사선산업학회지, 10(3), pp.139~144 (2016), (사)한국방사선산업학회[DOI]
  12. "GATE Model of a SPECT-CT Equi pment for Breast Cancer Diagnosis" (2016)[URL]
  13. Wang, W, "Validation of Monte Carlo simulation of mammography with TLDmeasurement and depth dose calculation with a detailed breast model", EPJ Web of Conferences, 153, pp.04017~ (2017)
  14. 홍동희, "유방촬영 시 피폭선량을 줄이기 위한 압박대 재질의 PC, PMMA, Carbon에 대한 연구", 방사선기술과학, 38(2), pp.81~87 (2015), 대한방사선과학회[DOI]
  15. 이인자, "초점", 필터에 따른 유방 압박 두께와 평균 유선 선량의 관계, 2009, pp.32(3), 261~270[DOI]