개발한 진단용 다엽조리개 성능평가 및 X선영상과 적외선체열영상의 융합영상 구현

Performance Evaluation of the Developed Diagnostic Multi-Leaf Collimator and Implementation of Fusion Image of X-ray Image and Infrared Thermography Image

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

We have developed and applied a diagnostic Multi-Leaf Collimator (MLC) to optimized the X-ray field in medical imaging and the usefulness evaluated through the fusion of infrared image and X-ray image acquired by infrared camera. The hand and skull radiography with multi-leaf collimator(MLC) showed significant area dose reductions of 22.9% and 31.3% compared to ARC and leakage dose was compliant with KS A 4732. Also scattering doses of 50 cm and 100 cm showed a significant decrease to confirm the usefulness of MLC. It was confirmed that the fusion of infrared images with an adjustable degree of transparency was possible in the X-ray images. Therefore, fusion of anatomical information with physiological convergence is expected to contribute and improvement of diagnostic ability. In addition, the feasibility of convergence X-ray imaging and DITI devices and the possibility of driving MLC with infrared images were confirmed.

Abstract

We have developed and applied a diagnostic Multi-Leaf Collimator (MLC) to optimized the X-ray field in medical imaging and the usefulness evaluated through the fusion of infrared image and X-ray image acquired by infrared camera. The hand and skull radiography with multi-leaf collimator(MLC) showed significant area dose reductions of 22.9% and 31.3% compared to ARC and leakage dose was compliant with KS A 4732. Also scattering doses of 50 cm and 100 cm showed a significant decrease to confirm the usefulness of MLC. It was confirmed that the fusion of infrared images with an adjustable degree of transparency was possible in the X-ray images. Therefore, fusion of anatomical information with physiological convergence is expected to contribute and improvement of diagnostic ability. In addition, the feasibility of convergence X-ray imaging and DITI devices and the possibility of driving MLC with infrared images were confirmed.

참고문헌 (22)

  1. Teng J, "A multimodality medical image fusion algorithm based on wavelet transform", International Conference in Swarm Intelligence, pp.627~633 (2010), Springer
  2. James AP, "Medical image fusion, A survey of the state of the art", Information Fusion, 19, pp.4~19 (2014)
  3. 정우경, "투시와 중재시술의 방사선 피폭과 저감화 방법", 대한의사협회지, 54(12), pp.1269~1276 (2011), 대한의사협회
  4. Korean society of medical imaging technology, "Textbook of radiographic positioning and clinical diagnosis" (2017), Chungku publisher
  5. Research society of radiation equipment, "Radiation equipment" (2017), Chungku publisher
  6. "KS A 4732. Variable X-ray beam limiting devices for diagnostic use", ~ (1997)
  7. "IEC 60601-1-33. General rule for X-ray equipment", ~ (2008)
  8. 한수철, "진단용 다엽콜리메이터 개발을 위한 몬테칼로 시뮬레이션 연구", 방사선기술과학, 39(4), pp.595~600 (2016), 대한방사선과학회[DOI]
  9. Seungwoo Park, "A Study on Development and Usefulness of new Multi-Leaf Collimator for Diagnostic X-ray Equipment", Journal of Electrical Engineering & Technology, 13(2), pp.936~942 (2018), 대한전기학회
  10. Le Heron JC, "Estimation of effective dose to the patient during medical X-ray examinations from measurements of the dose-area product", Physics in Medicine & Biology, 37(11), pp.2117~ (1992)
  11. 권순무, "디지털 유방촬영장치에서 다엽 조리개 적용으로 인한 화질과 선량의 변화", 한국방사선학회논문지, 9(7), pp.535~540 (2015), 한국방사선학회[DOI]
  12. 강영한, "디지털 심혈관조영장치의 기하학적 특성에 따른 선량 감소", 한국방사선학회논문지, 7(4), pp.277~284 (2013), 한국방사선학회[DOI]
  13. Bor D, "Comparison of effective doses obtained from dose–area product and air kerma measurements in interventional radiology", The British Journal of Radiology, 77(916), pp.315~322 (2004)
  14. 최성관, "X선 촬영에서 조사야 크기에 따른 산란선량의 변화", 한국콘텐츠학회 논문지, 13(3), pp.198~203 (2013), 한국콘텐츠학회[DOI]
  15. 조평곤, "투시 검사실 내 공간산란선 분포 측정", 한국콘텐츠학회 논문지, 11(10), pp.349~354 (2011), 한국콘텐츠학회
  16. Buchanan JB, "Tumor growth, doubling times, and the inability of the radiologist to diagnose certain cancers", Radiologic Clinics of North America, 21(1), pp.115~126 (1983)
  17. 박정규, "디지털 적외선 체열진단기를 이용한 추간판탈출증 환자의 ROI 온도측정", 한국방사선학회논문지, 11(4), pp.273~278 (2017), 한국방사선학회[DOI]
  18. Sim JH, "Diagnostic significance of digital infrared thermographic imaging after cerebrovascular stroke", Journal of Korean Neurosurgical Society, 27(7), pp.890~897 (1998)
  19. Sanchis-Sánchez E, "Infrared thermography is useful for ruling out fractures in pediatric emergencies", European Journal of Pediatrics, 174(4), pp.493~499 (2015)
  20. Kwon SM, "X-ray inspection device and its inspection method", Korea Patent 10-1743469, ~ (2017)
  21. Kwon SM, "Diagnosis method and device of breast disease", Korea Patent 10-1828367, ~ (2018)

관련 논문AI추천

이 간행물의 다른 논문 (7건)