장골 검출기를 이용한 장골 검사에서 3D 프린터로 제작한 구리 필터의 유용성 연구
A Study on the Usefulness of Copper Filters Made with 3D Printers in Longbone Examination Using Long Length Detector
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Long-bone examination is mainly used for inspection of the lower extremities. Recently, a long length detector (FXRD-1751S, VIEWORKS, Korea) with three digital detectors attached has been developed. High energy X-rays are used because pelvic areas require high image quality. In this case, X-rays are transmitted a lot in thin areas such as an ankle, and it is not suitable for diagnosing an image. Therefore, this study use copper filters made with 3D printers to increase image quality in the Long-bone inspection. A copper filter was manufactured in consideration of the overall thickness of the lower part. The experiment was conducted in anterior-posterior (AP) and lateral (LAT) positions, depending on the presence or absence of the filter. 5x5 pixels of region of interest (ROI) were selected from the pelvis, knee, and ankle areas. X-rays were irradiated under the conditions of 70 kVp and 40 mAs for AP, 80 kVp, and 63 mAs for lat when without filters, 90 kVp and 80 mAs for AP, 90 kVp and 100 mAs for lat when with filters. signal to noise ratio(SNR) ratio and contrast to noise (CNR) values were measured 1106.38, 14.34 before applying the filter and 1189.32, 70.43 after the filter. For the knee area, 650.44, 97.61 before applying the filter, and 1013.17, 444.24 after applying the filter. For the ankle area, 206.65, 23.68 before applying the filter and 993.50, 136.11 after applying the filter. In the Long-bone examination, SNR and CNR were greatly measured when the filter was applied, confirmed the availability of using the copper additional filter.
Abstract
Long-bone examination is mainly used for inspection of the lower extremities. Recently, a long length detector (FXRD-1751S, VIEWORKS, Korea) with three digital detectors attached has been developed. High energy X-rays are used because pelvic areas require high image quality. In this case, X-rays are transmitted a lot in thin areas such as an ankle, and it is not suitable for diagnosing an image. Therefore, this study use copper filters made with 3D printers to increase image quality in the Long-bone inspection. A copper filter was manufactured in consideration of the overall thickness of the lower part. The experiment was conducted in anterior-posterior (AP) and lateral (LAT) positions, depending on the presence or absence of the filter. 5x5 pixels of region of interest (ROI) were selected from the pelvis, knee, and ankle areas. X-rays were irradiated under the conditions of 70 kVp and 40 mAs for AP, 80 kVp, and 63 mAs for lat when without filters, 90 kVp and 80 mAs for AP, 90 kVp and 100 mAs for lat when with filters. signal to noise ratio(SNR) ratio and contrast to noise (CNR) values were measured 1106.38, 14.34 before applying the filter and 1189.32, 70.43 after the filter. For the knee area, 650.44, 97.61 before applying the filter, and 1013.17, 444.24 after applying the filter. For the ankle area, 206.65, 23.68 before applying the filter and 993.50, 136.11 after applying the filter. In the Long-bone examination, SNR and CNR were greatly measured when the filter was applied, confirmed the availability of using the copper additional filter.
참고문헌 (19)
- 김계선, "Digital Chest Tomosynthesis에서 부가필터에 따른 화질 및 유효선량", 방사선기술과학, 38(4), pp.347~353 (2015), 대한방사선과학회[DOI]
- 신경림, "도시지역 여성노인의 건강관심도, 주관적 건강평가, 건강상태 및 건강증진행위와의 관계연구", Journal of Korean Academy of Nursing, 34(5), pp.869~880 (2004), 한국간호과학회
- 김상현, "디지털 방사선장치에서 구리 부가필터의 유용성 평가에 관한 융복합 연구", 디지털융복합연구, 13(9), pp.351~359 (2015), 한국디지털정책학회[DOI]
- Cho BP, "Human Anatomy", pp.108~ (2019), Komoonsa
- Lee HS, "Consideration of the iliac measurement imaging method-The iliac measurement imaging method using a remote control X-ray device", Journal of the Korean Radiological Technologist Association, 24(2), pp.91~94 (1998)
- 장수한, "Longbone 검출기를 이용한 Scanogram의 유효성 평가", 방사선기술과학, 43(4), pp.235~242 (2020), 대한방사선과학회
- Jung PH, "A Study on Lower Extremity Scanogram Comparison according to the Examination Method", Journal of the Korean Society of Radiology, pp.13~19 (2016)
- Kang MT, "Patient dose due to use added filter during radiation examination", Journal of the Korean Society of Radiology, pp.65~169 (2015)
- Kim KM, "The Usefulness analysis of using additional filter in Pediatric chest X-ray", Journal of the Korean Society of Radiology, pp.189~194 (2017)
- Shin SI, "The Study on the Reduction of Patient Surface Dose Thorough the use of Copper Filter in a Digital Chest Radiography", Journal of Radiation Technology Science, 31(3), pp.223~228 (2008)
- 김상현, "복합부가여과에서 알루미늄 여과판 사용 시 양자검출효율과 화질평가를 통한 효율성 분석", 한국콘텐츠학회 논문지, 15(10), pp.362~373 (2015), 한국콘텐츠학회[DOI]
- 최일홍, "부가필터 사용 시 전방 산란선량에 따른 화질 영향에 대한 연구", 한국방사선학회논문지, 10(8), pp.597~602 (2016), 한국방사선학회[DOI]
- 장희민, "3차원 프린팅 기술을 이용한 차폐체 제작 및 유용성 평가", 한국방사선학회논문지, 13(7), pp.1015~1024 (2019), 한국방사선학회[DOI]
- 윤준, "FFF방식의 3D프린터 노즐 크기와 층 높이가 방사선 차폐체 제작에 미치는 영향에 관한 연구", 한국방사선학회논문지, 14(7), pp.891~898 (2020), 한국방사선학회[DOI]
- 윤명성, "3D프린터를 이용한 CT 선량측정 팬텀 제작 및 비교에 관한 연구", 한국방사선학회논문지, 12(6), pp.737~743 (2018), 한국방사선학회[DOI]
- 김영천, "하지전신계측검사에서 자세의 변화에 따른 방사선량 및 영상평가", 한국방사선학회논문지, 8(7), pp.409~415 (2014), 한국방사선학회[DOI]
- Jang SW, "The Usefulness analysis of Using Additional filter in Dual energy chest X-ray", Journal of the Korean Society of Radiology, pp.1~11 (2015)
- 최남길, "진단용 X-선 촬영시 부가 필터 및 노출의 변화에 따른 피폭선량 및 영상 화질 비교 연구", 방사선방어학회지, 37(1), pp.25~34 (2012), 테스트학회[DOI]
- 권경태, "단일조사 whole spine Lateral 검사에서3D 프린터로 제작한 구리 필터 유용성 연구", 한국방사선학회논문지, 14(7), pp.899~906 (2020), 한국방사선학회[DOI]