Geant4 코드를 이용한 선형가속기 6 MV 광자선의 선량분포에 관한 연구

Geant4 Code Based Simulation of 6 MV Photon Beam for Analysis of Dose Distribution

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study is to present a Geant4 code for the simulation of the absorbed dose distribution given by a medical linac for 6 MV photon beam. The dose distribution was verified by comparison with calculated beam data and beam data measured in water phantom. They were performed for percentage depth dose(PDD) and beam profile of cross-plane for two field sizes of 10 × 10 and 15 × 15 cm2. Deviations of a percentage and distance were obtained. In energy spectrum, the mean energy was 1.69 MeV. Results were in agreement with PDD and beam profile of the phantom with a tolerance limit. The differences in the central beam axis data δ1 for PDD had been less than 2% and in the build up region, these differences increased up to 4.40% for 10 cm square field. The maximum differences of δ2 for beam profile were calculated with a result of 4.35% and 5.32% for 10 cm, 15 cm square fields, respectively. It can be observed that the difference was below 4% in δ3 and δ4. For two field sizes of δ50-90 and RW50, the results agreed to within 2 mm. The results of the t-test showed that no statistically significant differences were found between the data for PDD of δ1, p>0.05. A significant difference on PDD was observed for field sizes of 10 × 10 cm2, p=0.041. No significant differences were found in the beam profile of δ3, δ4, RW50, and δ50-90. Significant differences on beam profile of δ2 were observed for field sizes of 10 × 10 cm2, p=0.025 and for 15 × 15 cm2, p=0.037. This work described the development and reproducibility of Geant4 code for verification of dose distribution.

Abstract

This study is to present a Geant4 code for the simulation of the absorbed dose distribution given by a medical linac for 6 MV photon beam. The dose distribution was verified by comparison with calculated beam data and beam data measured in water phantom. They were performed for percentage depth dose(PDD) and beam profile of cross-plane for two field sizes of 10 × 10 and 15 × 15 cm2. Deviations of a percentage and distance were obtained. In energy spectrum, the mean energy was 1.69 MeV. Results were in agreement with PDD and beam profile of the phantom with a tolerance limit. The differences in the central beam axis data δ1 for PDD had been less than 2% and in the build up region, these differences increased up to 4.40% for 10 cm square field. The maximum differences of δ2 for beam profile were calculated with a result of 4.35% and 5.32% for 10 cm, 15 cm square fields, respectively. It can be observed that the difference was below 4% in δ3 and δ4. For two field sizes of δ50-90 and RW50, the results agreed to within 2 mm. The results of the t-test showed that no statistically significant differences were found between the data for PDD of δ1, p>0.05. A significant difference on PDD was observed for field sizes of 10 × 10 cm2, p=0.041. No significant differences were found in the beam profile of δ3, δ4, RW50, and δ50-90. Significant differences on beam profile of δ2 were observed for field sizes of 10 × 10 cm2, p=0.025 and for 15 × 15 cm2, p=0.037. This work described the development and reproducibility of Geant4 code for verification of dose distribution.

참고문헌 (27)

  1. Ahnesjo A, "Dose calculations for external photon beams in radiotherap", Phys. Med. Biol, 44(11), pp.R99~R155 (1999)
  2. Verhaegen F, "Monte Carlo modelling of external radiotherapy photon beams", Phys. Med. Biol, 48(21), pp.R107~R164 (2003)
  3. Chauvie S, "Radiotherapy treatment planning with Monte Carlo on a distributed system", pp.1765~1769 (2004), IEEE
  4. Gonias P, "Monte Carlo Simulation of 6MV Varian Clinac Photon Beam Using GEANT4-GATE", J At Nucl Phys, 2(1), pp.30~38 (2020)
  5. 안수정, "GATE6를 이용한 Varian 21EX Clinac 선형가속기의 6 MV X-선 특성모사", 방사선기술과학, 39(4), pp.571~575 (2016), 대한방사선과학회[DOI]
  6. Agostinelli S, "Geant4-a simulation toolkit", Nuclear Instruments and Methods in Physics Research A, 506(3), pp.250~303 (2003)
  7. Sheikh-Bagheri D, "Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code", Med. Phys, 29(3), pp.391~402 (2002)
  8. Ivanchenko V, "Recent Improvements in Geant4 Electromagnetic Physics Models and Interfaces", Progress in Nuclear Science and Technology, 2, pp.898~903 (2011)
  9. Pandola L, "Validation of the GEANT4 simulation of bremsstrahlung from thick targets below 3 MeV", Nuclear Instruments and Methods in Physics Research B, pp.41~48 (2015)
  10. 강상구, "몬테카를로 방법에 의한 6MV 선형가속기의 광자 흡수선량 분포 평가에 관한 연구", 방사선기술과학, 34(1), pp.43~50 (2011), 대한방사선과학회
  11. Absorbed Dose Determination in External Beam Radiotherapy:An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, "International Atomic Energy Agency", ~, pp.Technical Reports Series No. 398 (2000)
  12. Commissioning and quality assurance of computerized planning systems for radiation treatment of cancer, "International Atomic Energy Agency", ~, pp.Technical Reports Series No.430 (2004)
  13. Venselaar J, "Tolerances for the accuracy of photon beam dose calculations of treatment planning systems", Radiotherapy and Oncology, 60, pp.191~201 (2001)
  14. Mesbahi A, "Monte Carlo calculation of Varian 2300C", D Linac photon beam characteristics: A comparison between MCNP4C, GEANT3 and measurements, 2005, pp.62, 469~477
  15. Baumgartner A, "Simulation of photon energy spectra from Varian 2100C and 2300C", D Linacs: Simplified estimates with PENELOPE Monte Carlo models, 2009, pp.67, 2007~2012
  16. Chetty IJ, "Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning", Med. Phys, 34(12), pp.4818~4836 (2007)
  17. Keall PJ, "Monte Carlo dose calculations for dynamic IMRT treatments", Phys. Med. Biol, 46 (2001)
  18. Ding GX, "Using Monte Carlo simulations to commission photon beam output factors-A feasibility study", Physics in Medicine & Biology, 48(23), pp.3865~3874 (2003)
  19. Mesbahi A, "Development and commissioning of a Monte Carlo photon beam model for Varian Clinac 2100EX linear accelerator", Applied Radiation and Isotopes, 64(6), pp.656~662 (2006)
  20. Fix MK, "Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy", Med Phys, 27(12), pp.2739~2747 (2000)
  21. Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code, "Sardari D", 64~68 (2010)
  22. Ding GX, "Dose discrepancies between Monte Carlo calculations and measurements in the buildup region for a high-energy photon beam", Med. Phys, 29(11), pp.2459~2463 (2002)
  23. Andreo P, "Monte Carlo techniques in medical radiation physics", Physics in Medicine & Biology, 36(7), pp.861~920 (1991)
  24. Bakkali JEL, "Validation of Monte Carlo Geant4 code for a 6 MV Varian linac", Journal of King Saud University-Science, 29, pp.106~113 (2017)
  25. JimRnez Spang F, "Monte Carlo Study of the Dosimetry of Small-Photon Beams Using CMOS Active Pixel Sensors", pp.2011
  26. Tzedakis A, "Influence of initial electron beam parameters on Monte Carlo calculated absorbed dose distributions for radiotherapy photon beams", Med. Phys, 31(4), pp.907~913 (2004)
  27. Cho SH, "Reference photon dosimetry data and reference phase space data for the 6 MV photon beam from Varian Clinac 2100 series linear accelerators", Med. Phys, 32(1), pp.137~148 (2005)

이 간행물의 다른 논문 (7건)