The Faraday cup electrode of different size has been developed and evaluated to investigate the diffusion effect of particles by Brownian motion in a particle beam mass spectrometer(PBMS). Particles which focused and accelerated by aerodynamic lens are charged to saturation in an electron beam, and then deflected electrostatically into a Faraday cup detector for measurement of the particle current. The concentration of particles is converted from currents detected by Faraday cup. Measurements of particle current as a function of deflection voltage are combined with measured relationships between particle velocity and diameter, charge and diameter, and mass and diameter, to determine the particle size distribution. The particle currents were measured using 5, 10, 20, 40 ㎜ sized Faraday cup that can be move to one direction by motion shaft. The current difference for each sizes as a function of position was compared to figure out diffusion effect during transport. Polystyrene latex(PSL) 100, 200 ㎚ sized standard particles were used for evaluation. The measurement using 5 ㎜ sized Faraday cup has the highest resolution in a diffusion distance and the smaller particles had widely diffused.
Abstract
The Faraday cup electrode of different size has been developed and evaluated to investigate the diffusion effect of particles by Brownian motion in a particle beam mass spectrometer(PBMS). Particles which focused and accelerated by aerodynamic lens are charged to saturation in an electron beam, and then deflected electrostatically into a Faraday cup detector for measurement of the particle current. The concentration of particles is converted from currents detected by Faraday cup. Measurements of particle current as a function of deflection voltage are combined with measured relationships between particle velocity and diameter, charge and diameter, and mass and diameter, to determine the particle size distribution. The particle currents were measured using 5, 10, 20, 40 ㎜ sized Faraday cup that can be move to one direction by motion shaft. The current difference for each sizes as a function of position was compared to figure out diffusion effect during transport. Polystyrene latex(PSL) 100, 200 ㎚ sized standard particles were used for evaluation. The measurement using 5 ㎜ sized Faraday cup has the highest resolution in a diffusion distance and the smaller particles had widely diffused.
참고문헌 (15)
- A. A. Scheidemann, "Faraday cup detector array with electronic multiplexing for multichannel mass spectrometry", J. Vac. Sci. Technol. A, 20, pp.597~604 (2002)
- A.K. Knight, "The development of a micro- Faraday array for ion detection", International Journal of Mass Spectrometry, 215, pp.131~139 (2002)
- Jeonggil Na, "Real-time diagnosis of nano-sized contaminant particles generated in TiN metal organic chemical vapor deposition", Applied Physics Express, 2, pp.035501~ (2009)
- Jeonggil Na, "Effects of process variables on TiN particles formation during metallorganic chemical vapor deposition", Electrochemical and Solid-State Letters, 13, pp.248~252 (2010)
- 나정길, "ISPM 및 PBMS를 이용한 BPSG 및 PSG CVD 공정 중발생하는 오염입자의 실시간 측정", PARTICLE AND AEROSOL RESEARCH, 6(3), pp.139~145 (2010), 한국입자에어로졸학회
- Paul J. Ziemann, "Electron impact charging properties of size-selected, submicrometer organic particles", J. Phys. Chem., 99, pp.5126~5138 (1995)
- Paul J. Ziemann, "Particle beam mass spectrometry of submicron particles charged to saturation in an electron beam", J. Aerosol Sci, 26(5), pp.745~756 (1995)
- Paul J. Ziemann, "Effects of particle shape and chemical composition on the electron impact charging properties of submicron inorganic particles", J. Aerosol Sci, 27(4), pp.587~606 (1996)
- Peng Liu, "Generating particle beams of controlled dimensions and divergence: Ⅰ. theory of particle motion in aerodynamic lenses and nozzle expansions", Aerosol Science and Technology, 22, pp.293~313 (1995)
- Peng Liu, "Generating particle beams of controlled dimensions and divergence: Ⅱ. experimental evaluation of particle motion in aerodynamic lenses and nozzle expansions", Aerosol Science and Technology, 22, pp.313~324 (1995)
- Robert B. Darling, "Micromachined Faraday cup array using deep reactive ion etching", Sensors and Actuators A, 95, pp.84~93 (2002)
- Sandeep Nijhawan, "Particle transport in a parallel-plate semiconductor reactor: Chamber modification and design criterion for enhanced process cleanliness", J. Vac. Sci. Technol. A, 18, pp.2198~2206 (2000)
- 김태성, "Particle formation during low-pressure chemical vapor deposition from silane and oxygen: Measurement, modeling, and film properties", JOURNAL OF VACUUM SCIENCE TECHNOLOGY A-VACUUM SURFACES AND FILMS, 20(2), pp.413~423, A V S AMER INST PHYSICS
- Taesung Kim, "A study of particle formation and transport during thermal chemical vapor deposition( CVD) of silicon dioxide films and high-density plasma CVD of poly silicon films", Ph. D, pp.2002
- William C. Hinds, "Aerosol technology: properties, behavior, and measurement of aerborne particls", pp.156~157 (1999), John Wiley & Sons, Inc