본문 바로가기
KERD
논문KCI 등재

SS-OCT용 파장 스위핑 레이저를 위한 스위핑 중심 세팅 자동화

Sweeping Center Setting Automation for Wavelength Swept Laser used in SS-OCT

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류공학일반

초록

In this paper, the automation of sweeping center setting for wavelength swept laser used in SS-OCT has implemented. For 3 regions where the initial FFP-TF pass wavelength can be located, each different DC voltage pattern is applied to FFP-TF. Through its per- formance test to the laser, fast and exact setting to sweeping central wavelength, flat sweeping with ±0.5 dB fluctuation range, and10 mW average optical power were obtained. This shows that the realized automatic setting process can replace an inconvenient manualsetting operation used for current wavelength swept laser. Additionally it cuts costs for optical spectrum analyzer necessary to laser spec- trum monitoring.

Abstract

In this paper, the automation of sweeping center setting for wavelength swept laser used in SS-OCT has implemented. For 3 regions where the initial FFP-TF pass wavelength can be located, each different DC voltage pattern is applied to FFP-TF. Through its per- formance test to the laser, fast and exact setting to sweeping central wavelength, flat sweeping with ±0.5 dB fluctuation range, and10 mW average optical power were obtained. This shows that the realized automatic setting process can replace an inconvenient manualsetting operation used for current wavelength swept laser. Additionally it cuts costs for optical spectrum analyzer necessary to laser spec- trum monitoring.

참고문헌 (13)

  1. Mark Brezinski, "Optical Coherence Tomography : Principles and Applications" (2006), Elsevier
  2. David Huang James G. Fujimoto, "Optical Coherence Tomography", Science, 254, pp.1178~1181 (1991)
  3. 김영관, "PZT 광경로 지연기를 이용한 시간영역 광간섭 단층촬영(OCT)에 관한 연구", 박사, 전자공학, pp.2011
  4. Y. Yasuno, "Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments", Opt. Express, 13, pp.10652~10664 (2005)
  5. R. Leitgeb, "Performance of Fourier domain vs. time domain optical coherence tomography", Opt. Express, 11, pp.889~894 (2003)
  6. M. A. Choma, "Sensitivity advantage of swept source and Fourier domain optical coherence tomography", Opt. Express, 11, pp.2183~2189 (2003)
  7. H. S. Kim, "Realization of Swept Source-Optical Coherence Tomography System using Loop Mirror within Reference Arm", COOC 2007, 14, pp.402~403 (2007)
  8. R. Huber, "Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging : design and scaling principles", Opt. Express, 13, pp.3513~3528 (2005)
  9. 엄진섭, "FDML 방식 스위핑 광원을 사용한 SS-OCT 구현", 센서학회지, 20(1), pp.46~52 (2011), 한국센서학회
  10. R. Huber, "Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography", Opt. Express, 14, pp.3225~3237 (2006)
  11. Eugene Hecht, "Optics" (1987), Addison-Wesley
  12. A. G. Podoleanu, "Unbalanced versus balanced operation in an optical coherence tomography system", Appl. Opt, 39, pp.173~182 (2000)
  13. 유욱재, "광섬유 pH 센서 개발을 위한 분광학적 연구", 센서학회지, 18(5), pp.365~371 (2009), 한국센서학회

관련 논문AI추천

이 간행물의 다른 논문 (11건)