Macular degeneration and glaucoma are representative age-related retinal diseases that rank second and third in the prevalence of retinal diseases, and are a kind of degenerative neurological disease. Irreversible visual acuity and visual field damage may occur, and the number of patients is rapidly increasing as the population ages. Since this retinal disease is a chronic disease, continuous drug treatment is required. There are various drug delivery methods for treatment, but direct injection of the drug into the intravitreal is the most effective for continuous delivery of the drug over a long period of time. In order to solidify Dexamethasone, a retinal disease treatment, and insert it into the primary intravitreal, it is important to develop a technology to miniaturize the treatment and an applicator to deliver the treatment. In this study, a mold technology was developed to integrate the cannula and hub, which are one part of applicator. In addition, surface treatment was performed on the outside of the cannula to improve the bonding strength between the cannula and the hub, and the bonding strength according to each condition was analyzed through a tensile test.
Abstract
Macular degeneration and glaucoma are representative age-related retinal diseases that rank second and third in the prevalence of retinal diseases, and are a kind of degenerative neurological disease. Irreversible visual acuity and visual field damage may occur, and the number of patients is rapidly increasing as the population ages. Since this retinal disease is a chronic disease, continuous drug treatment is required. There are various drug delivery methods for treatment, but direct injection of the drug into the intravitreal is the most effective for continuous delivery of the drug over a long period of time. In order to solidify Dexamethasone, a retinal disease treatment, and insert it into the primary intravitreal, it is important to develop a technology to miniaturize the treatment and an applicator to deliver the treatment. In this study, a mold technology was developed to integrate the cannula and hub, which are one part of applicator. In addition, surface treatment was performed on the outside of the cannula to improve the bonding strength between the cannula and the hub, and the bonding strength according to each condition was analyzed through a tensile test.
참고문헌 (9)
- 오민진, "고혈압성 안질환", 대한내과학회지, 83(5), pp.591~595 (2012), 대한내과학회[DOI]
- 류선영, "습성 나이관련황반변성의 유병률 변화와 라니비주맙 주사제 사용에 따른 향후 의료비용 분석", 대한안과학회지, 60(3), pp.253~260 (2019), 대한안과학회
- 김지훈, "탄소 섬유 강화 플라스틱과 금속의 접합에서 표면 패턴에 따른 접합 강도 영향", 한국생산제조학회지, 26(4), pp.430~435 (2017), 한국생산제조학회[DOI]
- 이민식, "Al5052", CFRP 복합소재의 표면특성이 접착성과 기계적특성에 미치는 영향, 2013, pp.26(5), 295~302[DOI]
- 이경엽, "알루미늄", CFRP 복합재의 접착강도 향상을 위한 정화방법에 관한 연구, 2002, pp.10(3), 130~135[DOI]
- 정용준, "AL6061과 PBT 재료의 인서트 사출공정조건에 따른접합강도 특성 분석", 한국기계가공학회지, 20(5), pp.135~141 (2021), 한국기계가공학회[DOI]
- A Study on the Effects in Adhesion of Liquid Silicone Rubbers in Mold Surface and PBT Insert at LSR, "PBT Multi component Injection Molding", Park, H. P., 2014, pp.508~509
- 이정원, "정밀 의료기기 부품 생산을 위한 사출금형의 가스벤트 제어에 관한 연구", Design & Manufacturing, 14(2), pp.34~41 (2020), 한국금형공학회[DOI]
- 이준훈, "공막 내 삽입한 생체분해성 고분자 매트릭스를 이용한 스테로이드 지속 방출 방법의 개발", 대한안과학회지, 46(11), pp.1894~1902 (2005), 대한안과학회