Nationwide Survey of Blood Culture Performance Regarding Skin Disinfection, Blood Collection and Laboratory Procedure
1
조회
0
다운로드
10
인용
0
참고문헌
키워드
초록
Background: Although many laboratories use automated blood culture systems, adequate skin disinfection and optimal blood volume are still critical for successful culture. The authors undertook a nationwide survey to understand the current situation and problems of blood culture in Korea. Methods: A survey of blood culture was performed in March and April 2010, including disinfectants, blood collection intervals, and recommended blood volumes. The laboratory physicians described the storage condition of culture bottles before delivery to the equipment. For quality control, the positive rate and skin contamination rate were enlisted. Results: Replies to the survey were collected from 74 Korean hospitals. Povidone iodine after either isopropyl alcohol or ethanol application was the most common way of skin disinfection. Sampling of a second set of cultures was performed simultaneously in 38% of hospitals and after a 30-min interval in 50%. The recommended blood volume was predominantly 10 mL (69%), whereas 20 mL only in 24%. The bottles were stored at 37oC in 23% and at room temperature in 16% before installation, whereas 57% placed directly in the equipment during the night shift. Positive rates ranged 8-10% at 32% of hospitals, 5-8% at 23%, and <5% at 12%. Skin contamination rates were 2-3% at 32%, 1-2% at 27%, and >3% at 13% of hospitals, respectively. Conclusion: Skin disinfection method was rather variable. Sampling interval, blood volume, and storage of bottles should be standardized. More than 10% of the hospitals are required for quality improvement in terms of positive rate and skin contamination rates
Abstract
Background: Although many laboratories use automated blood culture systems, adequate skin disinfection and optimal blood volume are still critical for successful culture. The authors undertook a nationwide survey to understand the current situation and problems of blood culture in Korea. Methods: A survey of blood culture was performed in March and April 2010, including disinfectants, blood collection intervals, and recommended blood volumes. The laboratory physicians described the storage condition of culture bottles before delivery to the equipment. For quality control, the positive rate and skin contamination rate were enlisted. Results: Replies to the survey were collected from 74 Korean hospitals. Povidone iodine after either isopropyl alcohol or ethanol application was the most common way of skin disinfection. Sampling of a second set of cultures was performed simultaneously in 38% of hospitals and after a 30-min interval in 50%. The recommended blood volume was predominantly 10 mL (69%), whereas 20 mL only in 24%. The bottles were stored at 37oC in 23% and at room temperature in 16% before installation, whereas 57% placed directly in the equipment during the night shift. Positive rates ranged 8-10% at 32% of hospitals, 5-8% at 23%, and <5% at 12%. Skin contamination rates were 2-3% at 32%, 1-2% at 27%, and >3% at 13% of hospitals, respectively. Conclusion: Skin disinfection method was rather variable. Sampling interval, blood volume, and storage of bottles should be standardized. More than 10% of the hospitals are required for quality improvement in terms of positive rate and skin contamination rates
참고문헌 (19)
- Lamy B, "What is the relevance of obtaining multiple blood samples for culture? A comprehensive model to optimize the strategy for diagnosing bacteremia", Clin Infect Dis, 35, pp.842~850 (2002)
- Mylotte JM, "Blood cultures: clinical aspects and controversies", Eur J Clin Microbiol Infect Dis, 19, pp.157~163 (2000)
- Clinical and Laboratory Standards Institute, "Principles and procedures for blood cultures. Approved guideline. Document M47-A", Clinical and Laboratory Standards Institute, ~ (2007)
- Li J, "Effects of volume and periodicity on blood cultures", J Clin Microbiol, 32, pp.2829~2831 (1994)
- Shafazand S, "Blood cultures in the critical care unit: improving utilization and yield", Chest, 122, pp.1727~1736 (2002)
- Reimer LG, "Update on detection of bacteremia and fungemia", Clin Microbiol Rev, 10, pp.444~465 (1997)
- Cockerill FR 3rd, "Optimal testing parameters for blood cultures", Clin Infect Dis, 38, pp.1724~1730 (2004)
- Chandrasekar PH, "Clinical issues of blood cultures", Arch Intern Med, 154, pp.841~849 (1994)
- Lemming L, "Bactec 9240 blood culture system: to preincubate at 35 degrees C or not", Clin Microbiol Infect, 10, pp.1089~1091 (2004)
- Riedel S, "Blood cultures: key elements for best practices and future directions", J Infect Chemother, 16, pp.301~316 (2010)
- Linder N, "Topical iodine-containing antiseptics and subclinical hypothyroidism in preterm infants", J Pediatr, 131, pp.434~439 (1997)
- Calfee DP, "Comparison of four antiseptic preparations for skin in the prevention of contamination of percutaneously drawn blood cultures: a randomized trial", J Clin Microbiol, 40, pp.1660~1665 (2002)
- Barenfanger J, "Comparison of chlorhexidine and tincture of iodine for skin antisepsis in preparation for blood sample collection", J Clin Microbiol, 42, pp.2216~2217 (2004)
- Malani A, "Review of clinical trials of skin antiseptic agents used to reduce blood culture contamination", Infect Control Hosp Epidemiol, 28, pp.892~895 (2007)
- Hall KK, "Updated review of blood culture contamination", Clin Microbiol Rev, 19, pp.788~802 (2006)
- Panlilio AL, "Infections and pseudoinfections due to povidoneiodine solution contaminated with Pseudomonas cepacia", Clin Infect Dis, 14, pp.1078~1083 (1992)
- Towns ML, "Guidelines on blood cultures", J Microbiol Immunol Infect, 43, pp.347~349 (2010)
- Connell TG, "How reliable is a negative blood culture result? Volume of blood submitted for culture in routine practice in a children's hospital", Pediatrics, 119, pp.891~896 (2007)
- Eskira S, "Reduction of blood culture contamination rate by an educational intervention", Clin Microbiol Infect, 12, pp.818~821 (2006)