Purpose: To determine the effective release time (RT) for listening to Korean sounds when wearing hearing aids, we evaluated Korean non-sense syllabic word recognition scores (NSRS) and the clarity of conversation sounds, the size of noise, and the overall preferences as a function of RT. Methods: Twenty adults and elderly people with mild to severe sensorineural hearing loss were participated as subjects. In the quiet and noise environment, NSRS and preferences were evaluated after setting the RT of the multi-channel hearing aids to 70, 200, and 600 ms. Presentation level of non-sense syllabic word was 45 dB HL and signal-to-noise ratio was 6 dB in noise environment. Results: NSRS was not different as a function of RT in quiet, but NSRS of 200 ms was higher than 70 ms and 600 ms in noise environment. However, clarity, loudness of noise, and overall preference did not show a difference in RT in a quiet and noisy environment. Conclusion: Through this study, it was confirmed that Korean NSRS can be affected by RT. Therefore, in order to develop a hearing aid fitting formula that is effective in recognition of Korean speech sound, it is necessary to consider the RT.

Abstract

Purpose: To determine the effective release time (RT) for listening to Korean sounds when wearing hearing aids, we evaluated Korean non-sense syllabic word recognition scores (NSRS) and the clarity of conversation sounds, the size of noise, and the overall preferences as a function of RT. Methods: Twenty adults and elderly people with mild to severe sensorineural hearing loss were participated as subjects. In the quiet and noise environment, NSRS and preferences were evaluated after setting the RT of the multi-channel hearing aids to 70, 200, and 600 ms. Presentation level of non-sense syllabic word was 45 dB HL and signal-to-noise ratio was 6 dB in noise environment. Results: NSRS was not different as a function of RT in quiet, but NSRS of 200 ms was higher than 70 ms and 600 ms in noise environment. However, clarity, loudness of noise, and overall preference did not show a difference in RT in a quiet and noisy environment. Conclusion: Through this study, it was confirmed that Korean NSRS can be affected by RT. Therefore, in order to develop a hearing aid fitting formula that is effective in recognition of Korean speech sound, it is necessary to consider the RT.

참고문헌 (29)

  1. Alexander, J. M., "Effects of WDRC release time and number of channels on output SNR and speech recognition", Ear and Hearing, 36(2), pp.e35~e49 (2015)
  2. American National Standards Institute, "ANSI S3.42-1992: Testing Hearing Aids with a Broad-Band Noise Signal", Acoustical Society of America, ~ (1992)
  3. American National Standards Institute, "ANSI S3.22-2003: Specification of Hearing Aid Characteristics", Acoustical Society of America, ~ (2003)
  4. Barker, C., "Client preferences for compression threshold in single-channel wide dynamic range compression hearing aids", Ear and Hearing, 20(2), pp.127~139 (1999)
  5. Barker, C., "Fitting low ratio compression to people with severe and profound hearing losses", Ear and Hearing, 22(2), pp.130~141 (2001)
  6. Boike, K. T., "Effect of compression ratio on speech recognition and speech-quality ratings with wide dynamic range compression amplification", Journal of Speech, Language, and Hearing Research, 43(2), pp.456~468 (2000)
  7. Byrne, D., "NAL-NL1procedure for fitting nonlinear hearing aids : Characteristics and comparisons with other procedures", Journal of the American Academy of Audiology, 12(1), pp.37~51 (2001)
  8. 최민정, "압축역치의 조절을 통한 음질의 개선 사례보고", Audiology and Speech Research, 7(1), pp.100~105 (2011), 한국청각언어재활학회
  9. Gatehouse, S., "Linear and nonlinear hearing aid fittings-1. Patterns of benefit", International Journal of Audiology, 45(3), pp.130~152 (2006)
  10. Gilbert, G., "Discrimination of release time constants in hearing-aid compressors", International Journal of Audiology, 47(4), pp.189~198 (2008)
  11. 한우재, "Preferred Compression Threshold and Release Time in Quiet and Noisy Conditions for Elderly Korean Hearing Aid Users", Journal of Audiology & Otology, 21(3), pp.133~139 (2017), 대한청각학회
  12. Hansen, M., "Effects of multi-channel compression time constants on subjectively perceived sound quality and speech intelligibility", Ear and Hearing, 23(4), pp.369~380 (2002)
  13. Henning, R. L., "The effects of hearing aid compression parameters on the short-term dynamic range of continuous speech", Journal of Speech, Language, and Hearing Research, 51(2), pp.471~484 (2008)
  14. Jenstad, L. M., "Quantifying the effect of compression hearing aid release time on speech acoustics and intelligibility", Journal of Speech, Language, and Hearing Research, 48(3), pp.651~667 (2005)
  15. Jin, I. K., "Derivations of the band-importance function: A cross-procedure comparison", The Journal of the Acoustical Society of America, 138(2), pp.938~941 (2015)
  16. 공혜경, "어음잡음하에서 자모음비에 따른 감각신경성난청인의 무의미음절인지도", Audiology and Speech Research, 9(1), pp.33~39 (2013), 한국청각언어재활학회
  17. 이경원, "한국 난청인을 위한 보청기 이득의 제안", Audiology and Speech Research, 7(2), pp.119~123 (2011), 한국청각언어재활학회
  18. 이경원, "한국 단음절어의 주파수중요기능 연구", Audiology and Speech Research, 8(1), pp.24~33 (2012), 한국청각언어재활학회
  19. Lee, K. W., "Comparing 2-cc coupler gain of monaural fitting with non-linear fitting formulas for elderly Korean with SNHL", Audiology and Speech Research, 4(1), pp.69~73 (2008)
  20. Lee, K. W., "Suggestions of multi-talker babble noise using Korean speech sound", Audiology and Speech Research, 4(1), pp.24~27 (2008)
  21. Lee, S. Y., "Case report for background noise reduction by compression threshold(CT)control in non-programmable hearing aids", Audiology and Speech Research, 4(2), pp.199~203 (2008)
  22. 이소예, "한국어의 자모음비(CVR)에 따른 무의미음절의 단어인지도 변화", Audiology and Speech Research, 6(1), pp.25~29 (2010), 한국청각언어재활학회
  23. Neuman, A. C., "Effect of release time in compression hearing aids : Paired-comparison judgments of quality", The Journal of the Acoustical Society of America, 98(6), pp.3182~3187 (1995)
  24. Neuman, A. C., "The effect of compression ratio and release time on the categorical rating of sound quality", The Journal of the Acoustical Society of America, 103(5 Pt 1), pp.2273~2281 (1998)
  25. 노복임, "경사형 및 수평형 난청유형에 따른 단음절 인지에 관한 비교연구", Audiology and Speech Research, 8(1), pp.78~86 (2012), 한국청각언어재활학회
  26. Oh, Y. J., "Acoustic comparisons of vowel and plosive productions between the normal and the hearing-impaired children", Speech Sciences, 7(2), pp.51~70 (2000)
  27. Pottackal Mathai, J., "Effect of hearing aid release time and presentation level on speech perception in noise in elderly individuals with hearing loss", European Archives of Oto-Rhino-Laryngology, 274(2), pp.671~677 (2017)
  28. Souza, P. E., "Effects of compression on speech acoustics, intelligibility, and sound quality", Trends in Amplification, 6(4), pp.131~165 (2002)
  29. Souza, P. E., "The contribution of amplitude envelope cues to sentence identification in young and aged listeners", Ear and Hearing, 22(2), pp.112~119 (2001)