하악구치부에서 Bicon 임플란트에 의해 지지되는 고정성 국소의치의 광탄성 응력분석

A PHOTOELASTIC STRESS ANALYSIS OF FIXED PARTIAL DENTURES WITH BICON IMPLANTS ON MANDIBULAR POSTERIOR AREA

0

조회

0

다운로드

0

인용

0

참고문헌

주제분류치의학

초록

Abstract

Statement of problem: Several prosthetic options are available for the restoration of multiple adjacent implants. A passively fitting prosthesis has been considered a prerequisite for the success and maintenance of osseointegration. Passivity is a particular concern with multiple implants because of documented inaccuracies in the casting and soldering process. One way to avoid this problem is to restore the implants individually, however, the restorations of individual adjacent impants requires careful adjustment of interproximal contacts. Purpose: The purpose of this study was to compare the stress distribution pattern and amount surrounding Bicon implants with individual crowns and splinted restorations. Material and method: A photoelastic model of a human partially edentulous left mandible with 3 Bicon implants(4×11 mm) was fabricated. For non-splinted restorations, individual crowns were fabricated on 3 abutments (4.0×6.5 ㎜, 0。, 2.0 ㎜post, Bicon Inc., Boston, USA) After the units were cemented, 4 levels of interproximal contact tightness were evaluated: open, ideal (8 ㎛shim stock drags without tearing), medium(40 ㎛), and heavy(80 ㎛). Splinted 3-unit fixed partial dentures were fabricated and cemented to the model. Changes in stress distribution under simulated non-loaded and loaded conditions(7.5, 15, 30 lb) were analyzed with a circular polaricope. Results: 1. Stresses were distributed around the entire body of fin in Bicon implants. 2. Splinted restorations were useful for distribution of stress around implants especially with higher loads. 3. By increasing the contact tightness between the individually restored three implants, the stress increased in the coronal portion of implants. Conclusions: Ideal adjustment of the contact tightness was important to reduce the stresses around individually restored Bicon implants.

참고문헌 (26)

  1. "Mechanical loading of Bra ¨nemark implantsaffects interfacial bone modeling andremodeling", 9, pp.345~60
  2. "Forces andmoment on Bra ¨nemark implants", 4, pp.241~7 (1989)
  3. "Bending overload and implant fracture:a retrospective clinical analysis", 10, pp.326~34 (1995)
  4. "The influence of controlledocclusal overload on peri-implant tissue", 13, pp.677~83 (1998)
  5. "The influence of controlled occlusaloverload on peri-implant tissue", 15, pp.425~31 (2000)
  6. "A finiteelement survey of eleven endosseousimplants J Prosthet Dent", 63(4), pp.457~65
  7. "Alternative materials for three endosseousimplantsJ Prosthet Dent 1989Jun", pp.717~22
  8. "Clinical rationale for tooth stabilizationand splinting", 43, pp.1~6 (1999)
  9. "Reactive force distributionsfor teeth when loaded singly andwhen used as fixed partial denture abutments", 42(4), pp.411~6
  10. "Influence of restoration type on stress distributionin bone around implants:a threedimensionalfinite element analysis", 13, pp.82~90 (1998)
  11. "A photoelastic study to determinethe location of the nonrigid connector inafive-unit intermediate abutment prosthesisJ Prosthet Dent 1987", pp.454~7
  12. "Restorative-based treatmentplaning:determinig adequate support forimplant-retained fixed restoration", 5, pp.179~84 (1996)
  13. "Fixture design and overload influencemarginal bone loss and fixture successin the Branemark system", 3, pp.104~11 (1992)
  14. "Analysisof load transfer and stress distribution byan implant-supported fixed partial denture", 75(3), pp.285~91 (1996)
  15. "Comparative accuracy of implant impressionprocedures", 12(2), pp.112~21 (1992)
  16. "An alternative tosplinting multiple implants: use of the ITI system", 7(2), pp.114~9 (1998)
  17. "The rotational aspect ofmandibular flexure", 64, pp.483~5 (1990)
  18. "Effect of splinting and interproximal contacttightness on load transfer by implantrestorations", 87(5), pp.528~35 (2002)
  19. "Deformation ofthe human mandible during simulatedtooth clenching", 73, pp.56~66 (1994)
  20. "Mandibular deformationin subjects with osseointegratedimplants", pp.319~28
  21. "Biomechanical concerns withfixed partial dentures involving implants", 2(4), pp.221~42
  22. "Finite elementstress analysis on the effect ofsplinting in fixed partial dentures", 81(6), pp.721~8 (1999)
  23. "Experimental stress analysis of dentalrestorations Structural design andstress analysis of fixed partial dentures", (abstract available), pp.177~86
  24. "The newconcept of ITI hollow -cylinder and hollowscrewimplants", pp.161~72
  25. "Comparison ofperi-implant stresses transmitted by fourcommercially available osseointegratedimplants", 9(3), pp.221~30 (1989)
  26. "The locking taperattachment for implant abutments: useand reliability", 5(4), pp.257~61 (1996)

관련 논문AI추천

하악 구치부에서 임플랜트 배열방식에 따른 임플랜트지지 고정성 국소의치의 광탄성 응력 분석김난영, 김유리, 조혜원대한치과보철학회 · 대한치과보철학회지2005 · 제43권 제1호연결고정, 인접면 접촉강도 및 치관길이에 따른 엔도포어 임플란트를 이용한 고정성 국소의치의 광탄성 응력 분석김유리, 정회열, 조혜원, 최민호대한치과보철학회 · 대한치과보철학회지2004 · 제42권 제4호상악 임플란트 overdenture에서 anchorage system과 의치상 구개피개가 하중전달에 미치는 영향임장섭, 전영찬, 정창모, 제홍지, 황재석대한치과보철학회 · 대한치과보철학회지2004 · 제42권 제4호구치부 부분 무치악 결손에서 유용한 세 가지 임프란트 수복법들의 응력분산에 관한 3차원 유한요소법적 비교연구김일규, 류승현, 변효인, 손충렬, 윤영아, 이성호, 최진호, 한예숙대한구강악안면외과학회 · 대한구강악안면외과학회지2004 · 제30권 제3호인공치아의 즉시부하를 위해 새로이 개발된 인공치아 지대주(Satellite Abutment)의 광탄성 응력 분석김부동, 박상규, 이백수, EngelkeW대한구강악안면외과학회 · 대한구강악안면외과학회지2002 · 제28권 제6호임플랜트 지지 보철물에서 고정체의 식립위치와 각도에 따른 삼차원 유한요소법적 응력분석에 관한 연구박원희, 이영수대한치과보철학회 · 대한치과보철학회지2005 · 제43권 제1호

이 간행물의 다른 논문 (9건)