본문 바로가기
KERD
논문KCI 등재

Quinine-Cd 착물을 Ionophore로 이용한 Iodide 이온 선택성 고체 접촉 전극

Iodide Ion Selective Solid Contact Electrode based on Quinine-Cd Ionophore

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Iodide ion selective poly(aniline) solid contact electrode based on quinine-Cd ionophore as a sensing material has been successfully developed. The electrode exhibits good linear response of 51.6 mV/decade (at 20±0.2℃, r²=0.9996) within the concentration range of 1×10^(-5.11) - 1×10^(-1.00) M KI. The composition of this electrode was quinine-Cd 3.0 : PVC 30.0 : bis(2-ethylhexyl)sebacate 67.0 (mass). This plasticizer provides the best response characteristics. The electrode shows good selectivity for iodide ion in comparison with any other anions and is suitable for use with aqueous solutions of pH 3.5 - 8.7. The standard deviations of the measured EMF difference were ± 1.52 and ± 1.73 mV for iodide sample solutions of 1.0×10^(-2) M and 1.0×10^(-4) M, respectively. The stabilization time was less than 400 s. and response time was less than 43 s.

Abstract

Iodide ion selective poly(aniline) solid contact electrode based on quinine-Cd ionophore as a sensing material has been successfully developed. The electrode exhibits good linear response of 51.6 mV/decade (at 20±0.2℃, r²=0.9996) within the concentration range of 1×10^(-5.11) - 1×10^(-1.00) M KI. The composition of this electrode was quinine-Cd 3.0 : PVC 30.0 : bis(2-ethylhexyl)sebacate 67.0 (mass). This plasticizer provides the best response characteristics. The electrode shows good selectivity for iodide ion in comparison with any other anions and is suitable for use with aqueous solutions of pH 3.5 - 8.7. The standard deviations of the measured EMF difference were ± 1.52 and ± 1.73 mV for iodide sample solutions of 1.0×10^(-2) M and 1.0×10^(-4) M, respectively. The stabilization time was less than 400 s. and response time was less than 43 s.

참고문헌 (25)

  1. N. N. Greenwood, , Chem. Elem, pp.933~ (1984)
  2. R. M. Calligaris, "Comprehensive Coordination Chemistry" (1987)
  3. Y. Yabu, , Endocrinol. Jpn, 33, pp.905~ (1986)
  4. A. S. Buinovskii, , Diagnostika Materialov, 70, pp.3~ (2004)
  5. M. Ying, , Analyst, 122, pp.1143~ (1997)
  6. F. El Aamrani, , Anal. Chim. Acta, 402, pp.129~ (1999)
  7. F. El Aamrani, , Anal. Chim. Acta, 329, pp.247~ (1996)
  8. S. Changqing, , Talanta, 46, pp.15~ (1998)
  9. Z. Li, , Anal. Lett, 30, pp.1455~ (1997)
  10. C. J. Coetzee, , Anal. Chem, 41, pp.1128~ (1969)
  11. Y. Umezawa, "CRC Handbook of Ion Selective Electrodes: Selectivity Coefficients" (1990), CRC Press
  12. V. J. Worting, , Anal. Chem, 62, pp.1506~ (1990)
  13. R. Yuan, , Anal. Chem, 65, pp.2572~ (1993)
  14. M. Shamsipur, , Polish J. Chem, 74, pp.231~ (2000)
  15. S. Daunert, , Anal. Chem, 61, pp.499~ (1989)
  16. V. J. Worting, , Anal. Chem, 62, pp.1506~ (1990)
  17. R. Yuan, , Anal. Chem, 65, pp.2572~ (1993)
  18. M. Shamsipur, , Polish J. Chem, 74, pp.231~ (2000)
  19. Jeong-Sik Yeom, , Journal of Electroanalytical Chemistry, 463, pp.16~ (1999)
  20. Won-Sik Han, , journal of Analytical chemistry, 66(9), pp.854~858 (2011)
  21. 홍태기, "메틸트리옥틸-암모늄 이온 Ionophore를 이용한 요오드 이온 선택성 고체 접촉 전극", 한국환경분석학회지, 13(3), pp.166~172 (2010), 한국환경분석학회
  22. M. Shamsipur, , Anal. Chimica Acta, 450, pp.37~ (2001)
  23. S. Daunert, , Anal. Chem, 61, pp.503~ (1989)
  24. L. S. Josefa, , Anal. Chim. Acta, 459, pp.229~ (2002)
  25. K. C. Oh, , J. Electroanal. Chem, 468, pp.98~ (1999)

이 간행물의 다른 논문 (10건)