초록

Real-time particle monitors are essential for accurately managing the air quality for fine particles and estimating exposure to the fine particles. However, many such monitors tend to be prohibitively expensive. A company (called as “A” hereafter) has recently developed a low-cost, real-time particle monitor, which is based on light scattering technique, to measure mass concentration of the fine particles in the air. In this study the “A” light scattering fine particle monitor was tested to provide reliability of the monitor both in ambient and indoor environments by comparing with filter-based (“gravimetric”) and another light scattering (GRIMM) methods. 24- hr integrated PM2.5 samples were collected at the ambient and indoor sites, and used to determine their mass and ionic species concentrations. Also black carbon (BC) concentration was determined using an Aethalometer. A measurement error of the monitor, which is defined as [(gravimetric method - “A” dust monitor)/gravimetric method × 100%], was examined with variations of BC and total secondary ionic species (SIC = NO3 -+SO4 2- +NH4 +) concentrations. In ambient environment, 24-hr average PM2.5 from the “A” light scattering dust monitor was approximately 13% higher than that from the gravimetric method with correlation coefficient (R2=0.92). Similar to the results from ambient measurements, the 24-hr average PM2.5 from the “A” dust monitor in an indoor environment of a building was strongly correlated with that from the gravimetric method (R2=0.96) with approximately 12% higher concentration in the “A” dust monitor. Also the PM2.5 mass concentration from the “A” dust monitor increased linearly (R2=0.84-0.98) with increasing the concentration of SIC which are known as light scattering components in the air. However, no clear association was found between the BC (light absorbing species) and measurement errors of the monitor, indicating no influence of BC particles to the response of the “A” light scattering dust monitor. Finally, results from this study indicates that the developed low-cost, realtime monitor using the light scattering technique could be useful for the fine particle mass determination in ambient and indoor environments. However, long-term measurements are required to better improve the accuracy and a linear relationship (“calibration factor”) of the dust monitor at a variety of sampling environments.

Abstract

Real-time particle monitors are essential for accurately managing the air quality for fine particles and estimating exposure to the fine particles. However, many such monitors tend to be prohibitively expensive. A company (called as “A” hereafter) has recently developed a low-cost, real-time particle monitor, which is based on light scattering technique, to measure mass concentration of the fine particles in the air. In this study the “A” light scattering fine particle monitor was tested to provide reliability of the monitor both in ambient and indoor environments by comparing with filter-based (“gravimetric”) and another light scattering (GRIMM) methods. 24- hr integrated PM2.5 samples were collected at the ambient and indoor sites, and used to determine their mass and ionic species concentrations. Also black carbon (BC) concentration was determined using an Aethalometer. A measurement error of the monitor, which is defined as [(gravimetric method - “A” dust monitor)/gravimetric method × 100%], was examined with variations of BC and total secondary ionic species (SIC = NO3 -+SO4 2- +NH4 +) concentrations. In ambient environment, 24-hr average PM2.5 from the “A” light scattering dust monitor was approximately 13% higher than that from the gravimetric method with correlation coefficient (R2=0.92). Similar to the results from ambient measurements, the 24-hr average PM2.5 from the “A” dust monitor in an indoor environment of a building was strongly correlated with that from the gravimetric method (R2=0.96) with approximately 12% higher concentration in the “A” dust monitor. Also the PM2.5 mass concentration from the “A” dust monitor increased linearly (R2=0.84-0.98) with increasing the concentration of SIC which are known as light scattering components in the air. However, no clear association was found between the BC (light absorbing species) and measurement errors of the monitor, indicating no influence of BC particles to the response of the “A” light scattering dust monitor. Finally, results from this study indicates that the developed low-cost, realtime monitor using the light scattering technique could be useful for the fine particle mass determination in ambient and indoor environments. However, long-term measurements are required to better improve the accuracy and a linear relationship (“calibration factor”) of the dust monitor at a variety of sampling environments.

참고문헌 (19)

  1. Chung, A., "Comparison of real-time instruments used to monitor airborne particulate matter", Journal of the Air Waste Management Association, 51, pp.109~120 (2001)
  2. McNamara, M. L., "Correction factor for continuous monitoring of wood smoke fine particulate matter", Aerosol and Air Quality Research, 11, pp.315~322 (2011)
  3. Wallace, L. A., "Validation of continuous particle monitors for personal, indoor, and outdoor exposures", Journal of Exposure Science and Environmental Epidemiology, 21, pp.49~64 (2011)
  4. 김정호, "중량법에 의한 광산란법 PM2.5 보정계수에 관한 연구", 한국도시환경학회지, 14(1), pp.41~47 (2014), 한국도시환경학회
  5. Heal, M. R., "Intercomparison of five PM10 monitoring devices and the implications for exposure measurement in epidemiological research", Journal of Environmental Monitoring, 2, pp.455~461 (2000)
  6. J. D. Yanosky, "A comparison of two direct-reading aerosol monitors with the federal method for PM2. 5 in indoor air", Atmospheric Environment, 36, pp.107~113 (2002)
  7. Hansen, A, "The Aethalometer-an instrument for the real-time measurement of optical absorption by aerosol particles", Science of the Total Environment, 36, pp.191~196 (1984)
  8. E. Weingartner, "Absorption of light by soot particles : determination of the absorption coefficient by means of aethalometers", Journal of Aerosol Science, 34, pp.1445~1463 (2003)
  9. W. Arnott, "Towards aerosol light absorption measurements with a 7-wavelength Aethalometer : Evaluation with a photoacoustic instrument and 3-wavelength nephelometer", Aerosol Science and Technology, 39, pp.17~29 (2005)
  10. T. Kirchstetter, "Controlled generation of black carbon particles from a diffusion flame and applications in evaluating BC measurements methods", Atmospheric Environment, 41, pp.1874~1888 (2007)
  11. A. Virkkula, "A simple procedure for correcting loading effects of Aethalometer", Journal of the Air & Waste Management Association, 57, pp.1214~1222 (2007)
  12. 박승식, "광주 도심지역에서 측정한 Aethalometer 검댕입자 자료의 보정", 한국대기환경학회지, 25(6), pp.571~578 (2009), 한국대기환경학회
  13. S. S. Park, "Measurement of real time black carbon for investigating spot loading effects of Aethalometer data", Atmospheric Environment, 44, pp.1449~1455 (2010)
  14. L. Drinovec1, "The ‘dualspot’Aethalometer : an improved measurement of aerosol black carbon with real-time loading compensation", Atmospheric Measurement Techniques, 8, pp.1965~1979 (2015)
  15. S. R. Won, "Characterization of PM2. 5and PM10 concentration distribution at public facilities in Korea", Journal of Korean Society for Indoor Environment, 9(3), pp.229~238 (2012)
  16. 환경부(Ministry of Environment), "Management guidelines for indoor air quality of public transport vehicles manufacturing and operating", ~
  17. L. Y. Chan, "Exposure level of carbon monoxide and respirable suspended particulate in public transportation modes while commuting in urban area of Guangzhou, China", Atmospheric Environment, 36, pp.5831~5840 (2002)
  18. M. Braniš, "Performance of a photometer DustTrak in various indoor and outdoor environments", Abstracts of the 2005 Evaluations and Assessment Conference, pp.535~ (2005)
  19. X. Querol, "Variability of levels and composition of PM10and PM2. 5 in the Barcelona metro system", Atmospheric Chemistry and Physics, 12, pp.5055~5076 (2012)