The development of a high efficient transcranial magnetic stimulation adopted real time- charging-discharging circuit

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

In this study, we have been proposed the new type of a transcranial magnetic stimulation adopted a variable voltage capacitor with Cockcroft-Walton circuit and constant-frequency current resonant half-bridge inverter. This a transcranial magnetic stimulation has some merits compared with the conventional one. First, it doesn't require the high voltage transformer. And second, it has less switching losses, compact size and capability in adjusting the transcranial magnetic stimulation output energy precisely. In this paper, we have performed the output characteristics of a transcranial magnetic stimulation system which is well known as magnetic stimulation. The tested results are described as a function of pulse repetition rate and switching numbers of the half-bridge inverter.

Abstract

In this study, we have been proposed the new type of a transcranial magnetic stimulation adopted a variable voltage capacitor with Cockcroft-Walton circuit and constant-frequency current resonant half-bridge inverter. This a transcranial magnetic stimulation has some merits compared with the conventional one. First, it doesn't require the high voltage transformer. And second, it has less switching losses, compact size and capability in adjusting the transcranial magnetic stimulation output energy precisely. In this paper, we have performed the output characteristics of a transcranial magnetic stimulation system which is well known as magnetic stimulation. The tested results are described as a function of pulse repetition rate and switching numbers of the half-bridge inverter.

참고문헌 (5)

  1. George MS, "Mechanisms and the current state of transcranial magnetic stimulation", CNS Spectrums, 8, pp.496~514 (2003)
  2. Barker AT, "Non-invasive magnetic stimulation of the human motor cortex", Lancet, 1, pp.1106~1107 (1985)
  3. Migita K, "Transcranial magnetic coil stimulation of motor cortex in patients with central pain", Neurosurgery, 36, pp.1037~1039 (1995)
  4. Rollnik JD, "Repetitive transcranial magnetic stimulation for the treatment of chronic pain a pilot study", Eur Neurol, 48, pp.6~10 (2002)
  5. Whi-Young Kim, "Development of Pocket Insertion style Magnetic Curer that Apply 2loop3 pulsed Variable Magnetic type Probe for Urinary Treatment", Journal of Information and Communication Convergence Engineering, 6(1), pp.73~79 (2008), 한국정보통신학회